Graham Field 760000-INS-LAB-RevA08 user manual Specifications, Storage, Pump Specification

Page 14

Storage

1.Lay the mattress out flat and upside down.

2.Roll from the head end toward the foot end.

3.Stretch the foot-end strap around the rolled mattress to prevent unrolling.

4.Store mattress and pump in a dust-free environment with no exposure to direct sunlight.

sCaution: Do not fold, crease or stack mattress.

 

Specifications

Pump

Specification

Power Supply

120V System: AC 120V 60 Hz

 

 

Note: see rating label on product

Fuse Rating

T1A, 250V

 

 

 

Cycle time

10, 15, 20 or 25 minutes

 

 

 

Air input

>8 LPM

Dimensions (L x W x H)

11.4" x 7.9" x 4.6" (29.1 x 20 x 11.7 cm)

Weight

5 lb (2.2 kg)

 

 

 

Environment

Temperature

Operation: 50°F to 104°F (10°C to 40°C)

 

 

Storage: 5°F to 122°F ( -15°C to 50°C)

 

 

Shipping: 5°F to 158°F ( -15°C to 70°C)

 

 

 

 

Humidity

Operation: 10% to 90% non-condensing

 

 

Storage: 10% to 90% non-condensing

 

 

Shipping: 10% to 90% non-condensing

 

 

 

Classification

Class II, Type BF, IPX0

 

 

Applied Part: Air Mattress

 

 

Not suitable for use in the presence of a

 

 

flammable anesthetic mixture (No AP or

 

 

APG protection)

 

 

 

14

760000-INS-LAB-RevA08

Image 14
Contents 760000 Contents Intended use of this device Important safety precautionsIntroduction INS-LAB-RevA08 Equipment Symbols Functional earth for UL only Class II for UL onlyPump front Product descriptionPump and mattress system Front panel third row therapy mode adjustment Front panel first row cycle time adjustmentFront panel second row pressure adjustment Pump rearFront panel fourth row Unpacking InstallationPump and Mattress Installation Operation Pressure adjustment Quick disconnectTransport cap MaintenanceCleaning CPR functionAir filter replacement General MaintenanceFuse replacement Storage SpecificationsPump Specification Warranty Mattress SpecificationINS-LAB-RevA08 GF Health Products, Inc