3M QuesTemp 32 user manual Placement for monitoring/testing, Sensors, Natural Wet Bulb Thermometer

Page 7

4

Placement for monitoring/testing

Placement

Placement for monitoring/testing

The QUESTemp should be placed at a height of 3.5 feet (1.1m) for standing individuals or 2 feet (.6m) for seated individuals. Tripod mounting is recommended to get the unit away from anything that might block radiant heat or airflow. A 1/4"x20 threaded bushing on the bottom of the instrument allows mounting to a standard photographic tripod. Do not stand close to the unit during sampling.

Make sure that the wet bulb reservoir is filled with distilled water and that the cotton wick is clean and fully wetted. After adding water or placing the unit in a new environment, allow ten minutes for the globe and wet bulb readings to stabilize.

A series of dashes appear in the display if one of the following occur:

The Heat Index or Humidex is outside of its allowable range

The temperature is outside of its allowable range

A temperature sensor has failed

Sensors

Natural Wet Bulb Thermometer

The natural wet bulb thermometer gives an indication of the effects of humidity on an individual. Relative humidity and air flow are taken into account by measuring the amount of evaporative cooling taking place at a thermometer covered with a moistened wick. The QUESTemp uses a cotton wick immersed into a reservoir containing distilled water. Ordinary tap water should not be used, as the contaminants that are left behind after evaporation will shorten the life of the wick and cause high readings. If the wick is discolored it should be replaced. To replace the wick, slide the old wick off the top of the sensor. Place a new wick over the sensor, making sure that the bottom of the wick is down in the reservoir.

Globe Thermometer

The globe thermometer gives an indication of the radiant heat exposure on an individual due to either direct light or hot objects in the environment. This is accomplished by placing a temperature sensor inside a blackened copper sphere and measuring the temperature rise. The WBGT index is based on the response of a 6 inch diameter globe. The QUESTemp uses a 2 inch diameter globe for a faster response time. The temperature of the 2 inch globe is correlated to match that of a 6 inch globe.

As an option, a sensor array with a 6 inch diameter globe is available.

Dry Bulb Thermometer

The dry bulb thermometer measures the ambient air temperature. This measurement is used in the outdoor WBGT calculation when a high solar radiant heat load may be present. The series of white plates surrounding the sensor shield it from radiant heat.

Image 7
Contents Market Leader Table of Contents Acgih Check Wick and Fill Natural Wet Bulb Up and RunningFilling wet bulb reservoir Heat Index / Humidex Wet Bulb Globe TemperatureMeasurements QUESTemp32Displayed Items Keypad OperationGlobe Thermometer Placement for monitoring/testingSensors Natural Wet Bulb ThermometerRemote, Sensors 2 Tri-Sensor Weighted AverageSensors 2 Relative Humidity Sensor Approved 9-Volt Batteries Power OptionsPower Options Alkaline Battery ReplacementBattery Pack NiMH Battery Pack Measurements SpecificationsNon-rechargeable battery Type Battery LifeAppendix a Appendix a Heat Exposures TablesUnited States Navy Wbgtf III Light Moderate Heavy Electrical power research institute EpriAppendix B Accessories Contacting 3M Detection Customer ServiceWarranty Service Department and Technical Support 1 800About Us