Essential Instructions (continued)

•  Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes. Connect all products to the proper electrical and pressure sources.

•  To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.

•  When replacement parts are required, ensure that qualified people use replacement parts specified by Rosemount. Unauthorized parts and procedures can affect the product’s performance and place the safe operation of your process at risk. Look alike substitutions may result in fire, electrical hazards, or improper operation.

•  Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.

Note

The Rosemount 6081 and all other wireless devices should be installed only after the 1420 Wireless Gateway has been installed and is functioning properly. Wireless devices should also be powered up in order of proximity from the 1420 Wireless Gateway, beginning with the closest. This will result in a simpler and faster network installation.

Note

Shipping considerations for wireless products (Power Modules):

The unit was shipped to you without the power module installed. Please remove the power modules from the unit prior to shipping.

Primary lithium power modules are regulated in transportation by the U. S. Department of Transportation, and are also covered by IATA (International Air Transport Association), ICAO (International Civil Aviation Organization), and ARD (European Ground Transportation of Dangerous Goods). It is the responsibility of the shipper to ensure compliance with these or any other local requirements. Please consult current regulations and requirements before shipping.

The power module with the wireless unit contains two “C” size primary lithium/thionyl chloride power sources.

Each power module contains approximately 5 grams in each pack. Under normal conditions, the power module materials are self-contained and are not reactive as long as the power modules and the pack integrity are maintained. Care should be taken to prevent thermal, electrical or mechanical damage. Contacts should be protected to prevent premature discharge.

Power module hazards remain when cells are discharged. Power modules should be stored in a clean and dry area. For maximum power module life, storage temperature should not exceed 30 °C.

WARNING

Use only with Rosemount Smart Power Module PN 701PBKKF

WARNING

Potential Electrostatic Hazard – The plastic antenna may present a potential electrostatic ignition hazard and must not be rubbed or cleaned with a dry cloth.

WARNING

Mechanical Spark Hazard – The 6081 enclosure is made of aluminum alloy and given a protective polyurethane paint finish. However, care should be taken to protect it from impact or abrasion if located in Zone 0.

This applies to any Intrinsically Safe installation, whether the Class, Division or Zone system is used.

2

Image 2
Contents Wireless pH/ORP Transmitter Essential Instructions Read this before proceedingEssential Instructions Quick Start Guide 6081 Wireless pH Transmitter Menu Tree for 6081 pH Wireless Transmitter Product Description General SpecificationsFunctional Specifications Hazardous Location Approvals PH Calibration Procedure Auto Buffer Calibration Sensor Wiring General InformationPower Module Installation / Replacement Mounting Mounting on a Flat SurfacePipe Mounting Pipe mounting kit PN 23820-00/01 accommodates 1-½ in. pipeFM is Installation Is Installation Is Installation Instruction Sheet 6081-P Emerson Process Management