Emerson 6081-P instruction sheet Power Module Installation / Replacement

Page 9

Instruction Sheet

6081-P

LIQ_MAN_ABR_6081-P

October 2013

 

 

Power Module Installation / Replacement

The expected life of the power module (PN 701PBKKF) is a minimum of four years at reference conditions. This section describes the procedure for initial installation and replacement of the power module (PN 701PBKKF). The new power module should be stored in a safe place with a controlled environment until the 6081 is ready for live operation. Note that the power module is packed separately from the 6081 wireless transmitter upon delivery and must be installed initially.

For initial installation or replacement of the power module, follow these steps using a Philips-head screwdriver:

1.Unscrew the two long machine screws to remove the rear cover of the 6081. Separate the rear cover from the central housing by manually prying the sections apart. Do not use screwdrivers or tools to separate these housing parts. The parts are sealed with an o-ring.

2.Before installation, note the safety warning, disposal instructions and part information on the connection side label of the power module.

3.With the 6081 front display section facing away from you, align the power module pack with the curved surface of the pack facing towards you and the small protruding connector facing away from you. Make sure to align the power module and its keyed connector with the connection receptacle in the middle of the instrument’s terminal block area.

4.With gentle pressure, insert the keyed connector on the power module into the receptacle (labeled Power Module Connection on the drawing). The power module seats in the connection receptacle with an o-ring.

5.Confirm that the power module is fully inserted in the receptacle and properly aligned with the surrounding terminal block.

6.Replace the rear cover of the 6081 with the two screws to secure it to the central housing. Tighten screws and verify operation. Correct installation the rear cover will ensure that the power module is properly secured to power the transmitter.

7.DO NOT RETURN SHIP THE USED POWER MODULE to Rosemount Analytical. Dispose of spent power modules as a hazardous material in accordance with government regulations.

WARNING

Use only with Rosemount SMART Power Module (PN 701PBKKF)

9

Image 9
Contents Essential Instructions Read this before proceeding Wireless pH/ORP TransmitterEssential Instructions Quick Start Guide 6081 Wireless pH Transmitter Menu Tree for 6081 pH Wireless Transmitter General Specifications Functional SpecificationsProduct Description Hazardous Location Approvals PH Calibration Procedure Auto Buffer Calibration General Information Sensor WiringPower Module Installation / Replacement Mounting on a Flat Surface MountingPipe mounting kit PN 23820-00/01 accommodates 1-½ in. pipe Pipe MountingFM is Installation Is Installation Is Installation Instruction Sheet 6081-P Emerson Process Management