Emerson 6081-P instruction sheet Hazardous Location Approvals

Page 6

6081-P

Instruction Sheet

October 2013

LIQ_MAN_ABR_6081-P

 

 

EEPROM Warning

Sense Line Open Warning Factory Cal Warning Keyboard Warning

Once a fault or warning is detected, the display will show a message describing the problem. Sensor Temperature Range: -10 to 150°C (PT100 and PT1000)

Display: 2-line, 16 character display supports display of pH and mV units. Display shows temperature.

Approvals:

RFI/EMI:

EN-61326

EN 301 489-1 V1.2 2002

EN 301 489-17: V1.4.1 2002

EN 60950-1: 2001

EN 300 328 V 1.6.1 (2004-11)

Hazardous Location Approvals

Intrinsic Safety:

Special Conditions of Use:

1.The 6081 enclosure is made of aluminum alloy and given a protective polyurethane paint finish; however, care should be taken to protect it from impact or abrasion if located in zone 0. This applies to any

Intrinsically Safe installation, whether the Class, Division or Zone system is used.

2.Potential Electrostatic Hazard – The plastic antenna may present a potential electrostatic ignition hazard and must not be rubbed or cleaned with a dry cloth.

Intrinsically Safe,

Class I, Division 1

Groups ABCD/T4

Ta = -20 °C to +65 °C – 1400322;

IP66

Class 1 Zone 0, AEx ia IIC T4

Ta = -20 °C to +65 °C – 1400322;

IP66

Complies with FM standards:

FM3600:1998

FM3610: 2010, ISA60079-0:2009, ISA60079-11:2009.

FM3611: 2004

FM3810: 2005 ANSI/IEC 60529:2004

Class I, Division 1, Groups A/B/C & D

Class II, Division 1, Groups E/F& G

Class III

T4 Tamb : -20 to +65 °C

Type 4x, IP66

CE 1180 II1G

Baseefa 10 ATEX 0149X

Ex ia IIC T4 Ga (-20 °C ≤ Ta ≤ +65 °C (source CE cert-6081 Baseefa) Complies with Standards: EN 60079-0:2009, EN 60079-11:2007

Non-Incendive:

Nonincendive, Class I,

Division 2, Groups ABCD/T4

Ta = -20˚C to +65 °C

Class I, Division 2,

Groups A, B, D

Dust Ignition Proof

Class II, Division 2,

Groups F & G

T4 Tamb : -20 to +65 °C

Environmental: The operating atmosphere of the transmitter must be consistent with the appropriate hazardous location certifications.

6

Image 6
Contents Wireless pH/ORP Transmitter Essential Instructions Read this before proceedingEssential Instructions Quick Start Guide 6081 Wireless pH Transmitter Menu Tree for 6081 pH Wireless Transmitter General Specifications Functional SpecificationsProduct Description Hazardous Location Approvals PH Calibration Procedure Auto Buffer Calibration Sensor Wiring General InformationPower Module Installation / Replacement Mounting Mounting on a Flat SurfacePipe Mounting Pipe mounting kit PN 23820-00/01 accommodates 1-½ in. pipeFM is Installation Is Installation Is Installation Instruction Sheet 6081-P Emerson Process Management