RUSKA 2470

Users Manual

Elastic Distortion of the Cylinder

As the pressure is increased within a piston pressure gauge, the resulting stress produces a temporary and reversible deformation of the cylinder. The net effect is a change in the effective area of the piston-cylinder combination. If the change in the area is a linear function of the applied pressure, the relationship may be described by the equation:

Ae = A0 (1 + b1P + b2 P2 )

Where:

Pis the nominal pressure

Ae

is the effective area at a pressure, P

A0

is the area of the piston-cylinder assembly at a reference pressure level

b1 & b2

are coefficients of elastic distortion which are determined experimentally

Gravity

Since pressure is defined as force per unit area, anything that changes the force applied to the piston of a piston pressure gauge also changes the pressure produced by that gauge. Therefore, the effects of gravity on the masses loaded on the piston must be considered. The gravity correction is usually very significant and must be used during calculations to achieve the advertised accuracy of the piston pressure gauge.

Confusion has resulted from the English System of units concerning the terms, mass and weight. The International System of units does not leave room for ambiguity and should be used whenever possible.

It is recognized that some facilities still operate under the English System of units. Therefore, this manual provides calibration data and calculation instructions in both the English and the International System of units.

Corrections for local gravity can vary by as much as 0.5% thus it is very important to have a reliable value for the local acceleration of gravity. A gravity survey with an uncertainty better than 0.00001 m/s2 is recommended.

Buoyant Effect of the Air

According to Archimedes's principle, the weight of a body in a fluid is diminished by an amount equal to the weight of the fluid displaced. The weight of an object (in air) that has had its mass corrected for the effects of local gravity is actually less than that corrected value indicates. This reduction in weight is equal to the weight of the quantity of air displaced by the object, or the volume of an object multiplied by the density of the air. But the volume of an irregular shaped object is difficult to compute from direct measurement. Buoyancy corrections are usually made by using the density of the material from which the object is made. If the value of mass is reported in units of apparent mass vs. brass standards rather than of true mass, the density of the brass standards must be used. Apparent mass is described as the value the mass appears to have, as determined in air having a density of 0.0012 g/cm³, against brass standards of a density of 8.4 g/cm³, whose coefficient of cubical expansion is 5.4 x 10-5/ ºC, and whose value is based on true mass in value (see reference 4).

2-4

Image 16
Contents Ruska Limited Warranty and Limitation of Liability Table of Contents Appendices List of Tables Ruska List of Figures Ruska How to Contact Fluke Safety InformationIntroduction Compressed GasSymbols Used in this Manual Heavy WeightsLubricants and Seals Oxygen CompatibilitySpecifications Low Range PistonAccuracy Humidity RangeGeneral Piston Pressure Gauge Considerations Types of Piston Pressure GaugesControlled Clearance Cylinder Calculations Measurement of Pressure with the Piston Pressure GaugeGravity Elastic Distortion of the CylinderBuoyant Effect of the Air Temperature Reference Plane of MeasurementsReference Plane Determination Is the density of the test media Crossfloating BibliographyGeneral Piston Pressure Gauge Considerations Ruska Description General InformationDescription of the Mass Set Description of the Gauge BaseDescription Ruska Installation IntroductionAlways remove the thermometer before shipping the gauge base Installation Ruska Precautions OperationLow Range Piston Assembly GeneralOperation 24-580 Bearing 2460-4-25 2460-5-6 Handling the Low Range Piston and Cylinder Step Low Range Piston and Cylinder Showing O-Ring Groove Mid Range Piston Assembly Section View, Mid Range Piston/Cylinder Thrust Cylinder Assembly Spacer Bearing 2460-70-2 2460-70-3 High Range Piston Assembly 11. Retaining Nut and Bearing12. Section View, High Range Piston/Cylinder 13. Parts Required for High Range Piston Operation Establishing Pressure Gauge PressuresAutomating the Calculations and Data Storage 16. Float PositionMaintenance of the Gauge LeaksRuska Piston / Cylinder Cleaning Instructions General Information and PreparationFunctional Testing of Piston/Cylinder Assemblies Cleaning the Low Range Piston/Cylinder AssemblyPiston / Cylinder Cleaning Instructions Preparations for Cleaning the Low Range Cylinder Preparing the Low Range Cleaning Tool Step Materials for Cleaning the Mid Range Piston/Cylinder Mid Range Piston/CylinderPreparing the Mid Range Cleaning Tool Cleaning the High Range Piston/Cylinder Assembly Materials for Cleaning the High Range Piston CylinderCleaning the High Range Piston/Cylinder Assembly Gmq33.bmp 13. Cleaning the High Range Cylinder 14. Drying the High Range Cylinder Explanation of Pressure Calculation Worksheet Temp. Coef. x actual temperature expected temperature For English UnitsK2 = 1/ g1 1 − ρa / ρb For SI UnitsAe t Explanation of Pressure Calculation Worksheet a Ruska Date Ruska Expected Temperature, t Ruska Equation A-4 Air Density Equation A-4 Air DensityNitrogen Density English Units 0 to 1000 Psig Nitrogen Density English Units 1,000 to 15,000 PsigNitrogen Density SI Units 0 to 6.9 MPa Nitrogen Density SI Units 6.9 MPa to 100 MPaZero Air Density SI Units 0 MPa to 20.7 MPa Helium Density SI Units 0 to 6.9 MPaConversion Factors Table B-1. Conversion Factors To Convert From Multiply ByRuska Glossary GlossaryDUT +INF, -INF Pressure Factor Sink Rate Ruska

2470 specifications

The Fluke 2470 is a versatile and highly regarded data acquisition system designed for a wide range of applications in electrical testing and measurement. Known for its precision and reliability, the Fluke 2470 is particularly popular among professionals who require accurate data for analysis and reporting.

One of the standout features of the Fluke 2470 is its high-performance data acquisition capabilities. It supports multiple channel inputs, allowing users to monitor and log various electrical parameters simultaneously. This multi-channel functionality is essential for complex testing scenarios where multiple variables must be tracked in real time.

The Fluke 2470 utilizes advanced measurement technologies, including a high-resolution analog-to-digital converter that ensures superior measurement accuracy. This technology is crucial for applications that demand precise readings, such as research and development, quality assurance, and compliance testing.

Additionally, the device is equipped with a variety of input types, including voltage, current, and temperature, making it suitable for a wide array of testing applications. The flexibility in input compatibility allows users to customize their measurements according to specific project requirements.

Another significant characteristic of the Fluke 2470 is its user-friendly interface and software integration. The accompanying software enables users to configure measurements easily, set up logging intervals, and analyze data visually. This intuitive setup minimizes the learning curve for new users and enhances productivity.

Data storage is another critical feature of the Fluke 2470. It provides ample internal memory, allowing users to record extensive datasets over long periods without the need for continuous monitoring. The stored data can be easily exported for further analysis, making it ideal for long-term testing projects.

Moreover, the Fluke 2470 excels in terms of portability and ruggedness. Its compact design is built to withstand demanding environments, ensuring that it can deliver accurate data in the field just as reliably as in the lab.

In conclusion, the Fluke 2470 combines high-performance data acquisition, multi-channel capabilities, advanced measurement technologies, and user-friendly features, making it a top choice for professionals in a variety of fields. Its robust design ensures that it meets the demands of rigorous testing while providing accurate and reliable data, crucial for informed decision-making.