Service Note 8904A-04

Solution/Action:

Configure the unit to output a ramp waveform from OUTPUT 1 at approximately 1 kHz and 1 Volt. Verify with an oscilloscope whether the ramp waveform is inverted (ramps down instead of up). If this behavior is unwanted then replace T1 with the part number noted above.

Image 2
Contents 8904A Multifunction Synthesizer Option 006/H16 R V I C E N O T EModification Recommended Parts RequiredSolution/Action