2.1Required Equipment
-One
-Two dc current meters with resolution to 1 mA and capable of the maximum current the dc power supply can supply. If available, set the current limit on the dc power supply to 100 mA.
Note: Some power supplies incorporate current meters which may be applicable to this test.
-
-Oscilloscope
2.2Power Supply Setup
1)Set the dc power supply to ±5 V.
2)Make sure the dc power supply is turned off before proceeding to the next step.
3)Connect the positive (+) terminal of the power supply to the positive (+) terminal of the current meter number 1.
4)Connect the negative (±) terminal of the current meter number 1 to the VCC+ of the EVM (J7).
5)Connect the common ground terminal of the power supply to the ground GND on the EVM (J9).
6)Connect the negative (±) terminal of the power supply to the negative (±) terminal of the second current meter.
7)Connect the positive (+) terminal of the current meter number 2 to the VCC± of the EVM (J8).
Figure 2±1. Power Supply Connection
CURRENT |
| CURRENT | ||
METER 2 |
| METER 1 | ||
| POWER SUPPLY |
| ||
+ | ±5V | GND | +5V | ± |
± |
| + | ||
| J8 VCC± |
| J9 GND |
|
| J7 VCC+ |
| J6 VOCM |
|
|
| EVM |
|
|
|
| THS4130 |
|
|
Figures are not drawn to scale.
Using the THS4130 EVM |