4 Operation

Model 3010MB

 

 

 

 

 

 

 

 

 

####% Span

Slope=#### ppm/s

When the Span value displayed on the screen is sufficiently stable, press Enter. (Generally, when the Span reading changes by 1 % or less of the full scale of the range being calibrated for a period of ten minutes it is sufficiently stable.) Once Enter is pressed, the Span reading changes to the correct value. The instrument then automatically enters the Analyze func- tion.

4.4.3Span Failure

The analyzer checks the output of the cell at the end of the span. If the raw output of the cell is less than 0.5 uA/ppm O2, the span will not be accepted. The analyzer will return to the previous calibration values, trigger the System Alarm, and display in the VFD:

Span Failed!!

This message will be shown for five seconds and the instrument shall return to the Analyze mode. In the upper right hand corner of the VFD display “FCAL” will be shown. This message flag will help the operator troubleshoot in case calibration was initiated remotely. To reset the alarm and the flag message, the unit must be turned off by cycling the standby key . It will not reset if the next span cycle is correct.

A trace cell is unlikely to fail span. As explained before, when the sensor reaches the end of its useful life, the zero offset begins to rise until the analyzer finds the zero unsatisfactory. Nevertheless, feeding the wrong span gas or electronics failure could set this feature off at the end of the span. Consider this before replacing the cell.

4.5The Alarms Function

The Model 3010MB is equipped with 2 fully adjustable concentration alarms and a system failure alarm. Each alarm has a relay with a set of form C contacts rated for 3 amperes resistive load at 250 V ac. See figure in chapter 3, Installation and/or the Interconnection Diagram included at the back of this manual for relay terminal connections.

The system failure alarm has a fixed configuration described in chapter

3Installation.

The concentration alarms can be configured from the front panel as either high or low alarms by the operator. The alarm modes can be set as

4-16: Part I

Teledyne Analytical Instruments

Image 46
Contents Model 3010MB Teledyne Analytical InstrumentsModel 3010MB Copyright 1999 Teledyne Analytical InstrumentsTable of Contents IiiSpecific Model Information Instrument Serial NumberOverview Typical ApplicationsMain Features of the Analyzer Model 3010MB Operational Theory InstallationIi Part OperationPart I Control Unit MaintenancePart I AppendixIv Part Model 3010MB Percent Paramagnetic Oxygen AnalyzerControl Unit Inner Control Panel Percent Paramagnetic Oxygen Analyzer Part I Control UnitIntroduction Model 3010MB Front of Unmounted Control UnitRecognizing Difference Between LCD VFDControl Unit Interface Panel Model 3010MB Rear PanelRemote Probe Analog OutputsRS-232 Port Network I/OIntroduction Model 3010MB Electronics and Signal Processing Operational Theory IntroductionBlock Diagram of the 3010TB CU Electronics Operational Theory Model 3010MBUnpacking the Control Unit Mounting the Control UnitInstallation Model 3010MB NPTElectrical Connections Required Front Door ClearancePrimary Input Power Connections Examples Analog Output ConnectionsRange Voltage Current mA Alarm Relays Connectors arePart RS-232 Signals RS-232 Sig RS-232 Pin PurposeCommand Description Parameter SettingFET Series Resistance Testing the System Operation Using the Data Entry and Function Buttons Operation Model 3010MBSystem Function Span ZeroPart Setting up an Auto-Cal Password Protection Entering the PasswordInstalling or Changing the Password Characters Available for Password Definition System Self-Diagnostic Test LogoutVersion Screen Showing Negative Oxygen ReadingsZero and Span Functions Auto Mode Zeroing Zero CalManual Mode Zeroing Cell FailureAuto Mode Spanning Span CalManual Mode Spanning Alarms Function Span FailurePart I Range Function Range LimitsSetting the Analog Output Ranges Analyze Function Operation Model 3010MB Fixed Range AnalysisSignal Output Voltage Signal Current Signal Ppm O Output V dc Output mA dcOperation Model 3010MB Fuse Replacement Part I Control Unit MaintenancePower System Self Diagnostic TestAnalog Major Internal Components PreampPart Model 3010M Ii Part Percent Paramagnetic Oxygen Analyzer Operational Theory Precise Paramagnetic SensorPrinciples of Operation Operational Theory Model 3010M Cross Interference Sensor Rear view shownTotal Cross Interference Percent Paramagnetic Oxygen Analyzer Maintenance Routine MaintenanceSensor Replacement Output Goes NegativeMaintenance Model 3010M Operating Temperature 0-45 C SpecificationsPercent Paramagnetic Oxygen Analyzer Appendix Appendix Model 3010M Recommended 2-Year Spare Parts List Drawing ListAppendix Model 3010M Percent Paramagnetic Oxygen Analyzer Appendix Zero Cal ZeroSettlingMan ENT To Begin

3010MB specifications

The Teledyne 3010MB is a cutting-edge instrument designed for the analysis of various gases in a multitude of applications. It is renowned for its reliability, accuracy, and advanced features that cater to diverse industrial needs. This sophisticated gas analyzer is particularly favored in environmental monitoring, process control, and emissions measurement sectors.

One of the main features of the Teledyne 3010MB is its ability to measure gaseous components such as carbon dioxide, methane, and various other hydrocarbons. Its robust design is complemented by state-of-the-art technology, ensuring consistent performance even in challenging environments. The analyzer leverages a microprocessor-controlled system, which allows for real-time data processing and enhanced operational efficiency.

The Teledyne 3010MB employs advanced infrared (IR) measurement techniques, providing superior sensitivity and specificity compared to conventional methods. The optical system utilizes a multi-path sample cell, which significantly increases the interaction of light with the sample gas, thereby improving the detection limits. This feature is particularly useful when dealing with trace gas measurements, where minute concentrations need to be accurately quantified.

Another key characteristic of the 3010MB is its user-friendly interface, which facilitates ease of operation. The large, graphic LCD display provides clear readouts of measured concentrations, operational status, and diagnostics, enabling operators to quickly interpret data. Moreover, intuitive menus and controls streamline navigation, reducing the learning curve for new users.

For data management, the Teledyne 3010MB offers comprehensive connectivity options. It can interface with various data logging systems and can be integrated into existing monitoring networks. This adaptability enhances its functionality, allowing users to easily collect and analyze data over time.

Furthermore, the device is built to withstand harsh operating conditions, featuring durable materials and a design that minimizes maintenance requirements. Its compact footprint makes it suitable for installation in limited spaces, while its versatility accommodates both laboratory and field applications.

In summary, the Teledyne 3010MB stands out as a leading gas analyzer, boasting features such as precise infrared measurement technology, a user-centric interface, and robust data connectivity capabilities. Its reliability and accuracy make it an invaluable tool for professionals in environmental monitoring and industrial process control, solidifying its reputation as a premier choice in gas analysis.