Page 45

Percent Paramagnetic Oxygen Analyzer

Part I: Control Unit

 

 

 

 

 

 

 

 

 

####

ppm

Span

Slope=####

ppm/s

 

The beginning span value is shown in the upper left corner of the

display. As the span reading settles, the screen displays and updates informa- tion on Slope. Spanning automatically ends when the span output corre-

sponds, within tolerance, to the value of the span gas concentration. Then the instrument automatically returns to the analyze mode.

4.4.2.2Manual Mode Spanning

Press Span to start the Span function. The screen that appears allows you to select whether the span calibration is to be performed automatically or manually.

Span: Settling:MAN

<ENT> For Next

Use the Δ∇ keys to toggle between AUTO and MAN span settling. Stop when MAN appears, blinking, on the display. Press Enter to move to the next screen.

Press Enter to move to the next screen.

Calib.Holdingtime

Cal hold: 5 min

This menue allows the operator to set the time the analyzer should be held in the auto span mode. It does not affect anything in the Manual Mode. Just press Enter to continue.

Span Val: 000008.00 <ENT>Span <UP>Mod #

Press Δ (<UP>) to permit modification (Mod #) of span value.

Use the arrow keys to enter the oxygen concentration of the span gas you are using (209000.00 if you are using air). The < > arrows choose the digit, and the Δ∇ arrows choose the value of the digit.

Press Enter to enter the span value into the system and begin the span calibration.

Once the span has begun, the microprocessor samples the output at a predetermined rate. It calculates the difference between successive samplings and displays this difference as Slope on the screen. It takes several seconds for the first Slope value to display. Slope indicates rate of change of the Span reading. It is a sensitive indicator of stability.

Teledyne Analytical Instruments

Part I: 4-15

Image 45
Contents Teledyne Analytical Instruments Model 3010MBCopyright 1999 Teledyne Analytical Instruments Model 3010MBIii Table of ContentsInstrument Serial Number Specific Model InformationTypical Applications OverviewMain Features of the Analyzer Model 3010MB Ii Part InstallationOperational Theory OperationPart I MaintenancePart I Control Unit AppendixModel 3010MB Percent Paramagnetic Oxygen Analyzer Iv PartPercent Paramagnetic Oxygen Analyzer Part I Control Unit Control Unit Inner Control PanelFront of Unmounted Control Unit Introduction Model 3010MBVFD Recognizing Difference Between LCDModel 3010MB Rear Panel Control Unit Interface PanelRS-232 Port Analog OutputsRemote Probe Network I/OIntroduction Model 3010MB Operational Theory Introduction Electronics and Signal ProcessingOperational Theory Model 3010MB Block Diagram of the 3010TB CU ElectronicsMounting the Control Unit Unpacking the Control UnitNPT Installation Model 3010MBRequired Front Door Clearance Electrical ConnectionsPrimary Input Power Connections Analog Output Connections ExamplesRange Voltage Current mA Connectors are Alarm RelaysPart RS-232 Sig RS-232 Pin Purpose RS-232 SignalsParameter Setting Command DescriptionFET Series Resistance Testing the System Operation Operation Model 3010MB Using the Data Entry and Function ButtonsSpan Zero System FunctionPart Setting up an Auto-Cal Entering the Password Password ProtectionInstalling or Changing the Password Characters Available for Password Definition Logout System Self-Diagnostic TestShowing Negative Oxygen Readings Version ScreenZero and Span Functions Zero Cal Auto Mode ZeroingCell Failure Manual Mode ZeroingSpan Cal Auto Mode SpanningManual Mode Spanning Span Failure Alarms FunctionPart I Range Limits Range FunctionSetting the Analog Output Ranges Operation Model 3010MB Fixed Range Analysis Analyze FunctionVoltage Signal Current Signal Ppm O Output V dc Output mA dc Signal OutputOperation Model 3010MB Part I Control Unit Maintenance Fuse ReplacementSystem Self Diagnostic Test PowerAnalog Preamp Major Internal ComponentsPart Model 3010M Ii Part Precise Paramagnetic Sensor Percent Paramagnetic Oxygen Analyzer Operational TheoryPrinciples of Operation Operational Theory Model 3010M Sensor Rear view shown Cross InterferenceTotal Cross Interference Sensor Replacement Routine MaintenancePercent Paramagnetic Oxygen Analyzer Maintenance Output Goes NegativeMaintenance Model 3010M Specifications Operating Temperature 0-45 CPercent Paramagnetic Oxygen Analyzer Appendix Appendix Model 3010M Drawing List Recommended 2-Year Spare Parts ListAppendix Model 3010M Percent Paramagnetic Oxygen Analyzer Appendix Zero Cal ZeroSettlingMan ENT To Begin

3010MB specifications

The Teledyne 3010MB is a cutting-edge instrument designed for the analysis of various gases in a multitude of applications. It is renowned for its reliability, accuracy, and advanced features that cater to diverse industrial needs. This sophisticated gas analyzer is particularly favored in environmental monitoring, process control, and emissions measurement sectors.

One of the main features of the Teledyne 3010MB is its ability to measure gaseous components such as carbon dioxide, methane, and various other hydrocarbons. Its robust design is complemented by state-of-the-art technology, ensuring consistent performance even in challenging environments. The analyzer leverages a microprocessor-controlled system, which allows for real-time data processing and enhanced operational efficiency.

The Teledyne 3010MB employs advanced infrared (IR) measurement techniques, providing superior sensitivity and specificity compared to conventional methods. The optical system utilizes a multi-path sample cell, which significantly increases the interaction of light with the sample gas, thereby improving the detection limits. This feature is particularly useful when dealing with trace gas measurements, where minute concentrations need to be accurately quantified.

Another key characteristic of the 3010MB is its user-friendly interface, which facilitates ease of operation. The large, graphic LCD display provides clear readouts of measured concentrations, operational status, and diagnostics, enabling operators to quickly interpret data. Moreover, intuitive menus and controls streamline navigation, reducing the learning curve for new users.

For data management, the Teledyne 3010MB offers comprehensive connectivity options. It can interface with various data logging systems and can be integrated into existing monitoring networks. This adaptability enhances its functionality, allowing users to easily collect and analyze data over time.

Furthermore, the device is built to withstand harsh operating conditions, featuring durable materials and a design that minimizes maintenance requirements. Its compact footprint makes it suitable for installation in limited spaces, while its versatility accommodates both laboratory and field applications.

In summary, the Teledyne 3010MB stands out as a leading gas analyzer, boasting features such as precise infrared measurement technology, a user-centric interface, and robust data connectivity capabilities. Its reliability and accuracy make it an invaluable tool for professionals in environmental monitoring and industrial process control, solidifying its reputation as a premier choice in gas analysis.