5. Technical Data
Product Description: | Digital Ear Thermometer |
|
|
|
|
|
Model: | Gentle Temp 521 | |||||
Sensing Unit: | Thermopile |
|
|
|
|
|
Temperature Display: | ||||||
| ||||||
Measurement Accuracy: | Ear Measurement Mode ± 0.2°C (± 0.4°F) within 35.5°C to 42.0°C (95.9°F | |||||
|
| to 107.6°F), ± 0.3°C (± 0.5°F) for other range | ||||
| Surface Mode | ±0.3°C (± 0.5°F) within 22.0°C to 42.2°C (71.6°F | ||||
|
| to 108.0°F), for other range it is ±2°C (± 3.6°F) or | ||||
|
| 4% whichever is greater. | ||||
Measurement Range: | Ear Measurement Mode 34.0°C (93.2°F ) to 42.2°C (108.0°F) | |||||
| Surface Mode | |||||
Power Supply: | 3.0V DC, 1 CR2032 Lithium Button Battery | |||||
Power Consumption: | 0.015 W |
|
|
|
|
|
Battery Life: | With a new battery approx. 2,500 measurements or more (Ambient | |||||
| environment 25±15°C, 50±40%RH) | |||||
Operating Environment | 10°C (50°F) to 40°C (104°F), 0 < RH < 85% | |||||
Temp and Humidity: |
|
|
|
|
|
|
Storage Environment | < | RH | < | 85% | ||
Temp and Humidity: |
|
|
|
|
|
|
Protection against electric | Internally powered ME equipment |
16 shock: