Invacare 3650 manual Tub Operation, Water Temperature Setting

Page 25

SECTION 3—OPERATING INSTRUCTIONS

Tub Operation

WARNING

DO NOT stand in tub. Tub surfaces are slippery when wet. Serious injury may occur.

Before immersing the resident or operating the shower/massage wand, check that the water temperature is between 95° F and 105° F (35° C and 41° C) to avoid scalding the resident.

Disinfect the tub after each use to avoid resident contamination.

Water Temperature Setting

WARNING

ALWAYS hand test the incoming water temperature and the water temperature in the tub before immersing the resident. Failure to do so may result in a scalding injury.

The causes, symptoms, and effects of hyperthermia may be described as follows: Hyperthermia occurs when the internal temperature of the body reaches a level several degrees above the normal body temperature of 98.6° F. The symptoms of hyperthermia include an increase in the internal temperature of the body, dizziness, lethargy, drowsiness, and fainting. The effects of hyperthermia include:

•Failure to perceive heat

•Failure to recognize the need to exit the spa or hot tub •Unawareness of impending hazard

•Fetal damage in pregnant women •Phyical inability to exit the spa or hot tub •Unconsciousness resulting in the danger of drowning

The use of alcohol, drugs, or medication can greatly increase the risk of fatal hyper- thermia.

NOTE: For this procedure, refer to FIGURE 3.3.

Flow Water Thermometer ‐ measures the

Flow Water Thermometer

temperature of water coming from the tub

 

fill spout or shower/massage wand.

 

Water Temperature Control Valve ‐ sets and maintains constant temperature for all incoming water.

Water Temperature Valve

FIGURE 3.3 Water Temperature Setting

Part No 1118387

25

Side Entry Whirlpool Tub

Image 25
Contents Side Entry Whirlpool Tub Accessories Warning Table of Contents Limited Warranty Special Notes MaintenanceCabinet Door 3600XL Label LocationModel Typical Product ParametersModels 3600 and 3600XL FeaturesModel Installation Important Safety InstructionsGeneral Guidelines Plumbing ElectricalDisinfecting Maintenance and InspectionGeneral Guidelines About Installation Parts InstallationParts Information Preparing the Work AreaPreparing the Work Area Securing Tub Securing TubElectrical and Plumbing Hook-up Locations on the Tub Electrical InstallationParts and Installation Electrical Installation Plumbing Installation Plumbing Installation Post Installation Checklist Overview Operating InstructionsDisinfecting the Tub Using the Closed Loop System Opening the Tub Door Opening/Closing the Tub DoorClosing the Tub Door Tub Operation Water Temperature SettingFilling Tub Operating Shower/Massage WandFilling Tub HydromassageSafety Inspection Checklists Safety INSPECTION/ TroubleshootingCleaning the Door Seal Gasket CleaningCleaning Your Tub Removing Soap Film and Hard Water DepositsRequired OZ Heavy Duty Cleaning ProcedureCalibrating the Disinfectant Flow Meter Safety INSPECTION/TROUBLESHOOTING Symptoms Faults Solution TroubleshootingMaintenance Record Side Entry Whirlpool Tub USA Limited Warranty
Related manuals
Manual 36 pages 42.54 Kb