Guidance and manufacturer’s
The S8 device is intended for use in the electromagnetic environment specified below. The customer or the user of the S8 device should assure that the device is used in such an environment.
Immunity test | Compliance level | Electromagnetic | |||
level | |||||
|
|
|
| ||
|
|
|
| ||
Electrostatic | ±6 kV contact | ±6 kV contact | Floors should be wood, concrete or ceramic tile. If | ||
discharge (ESD) | ±8 kV air | ±8 kV air | floors are covered with synthetic material, the relative | ||
IEC |
|
| humidity should be at least 30%. | ||
|
|
|
| ||
Electrical fast | ±2 kV for power | ±2 kV | Mains power quality should be that of a typical | ||
transient/burst | supply lines |
| commercial or hospital environment. | ||
IEC | ±1 kV for input/ | Not Applicable |
|
| |
| output lines |
|
|
| |
|
|
|
| ||
Surge | ±1 kV differential | ±1 kV differential | Mains power quality should be that of a typical | ||
IEC | mode | mode | commercial or hospital environment. | ||
| ±2 kV common | ±2 kV common |
|
| |
| mode | mode |
|
| |
|
|
|
| ||
Voltage dips, short | <5% Ut (>95% dip in | < 12V (>95% dip in | Mains power quality should be that of a typical | ||
interruptions and | Ut) for 0.5 cycle | 240V) for 0.5 cycle | commercial or hospital environment. | ||
voltage variations | 40% Ut (60% dip in | 96V (60% dip in | If the user of the S8 device requires continued | ||
on power supply | Ut) for 5 cycles | 240V) for 5 cycles | operation during power mains interruptions, it is | ||
input lines. | 70% Ut (30% dip in | 168V (30% dip in | recommended that the S8 device be powered from an | ||
IEC | Ut) for 25 cycles | 240V) for 25 cycles | uninterruptible power source. |
| |
| <5% Ut (>95%dip in | <12V (>95%dip in |
|
| |
| Ut) for 5 sec | 240V) for 5 sec |
|
| |
|
|
|
| ||
Power frequency | 3 A/m | 3 A/m | Power frequency magnetic fields should be at levels | ||
(50/60Hz) magnetic |
|
| characteristic of a typical location in a typical | ||
field IEC |
|
| commercial or hospital environment. | ||
|
|
|
| ||
|
|
| Portable and mobile RF communications equipment | ||
|
|
| should be used no closer to any part of the S8 device, | ||
|
|
| including cables, than the recommended separation | ||
|
|
| distance calculated from the equation applicable to | ||
|
|
| the frequency of the transmitter. | ||
|
|
| Recommended separation distance | ||
Conducted RF | 3 Vrms | 3 Vrms | d = 1.17 √P |
| |
IEC | 150 kHz to 80 MHz |
|
|
| |
Radiated RF | 10 V/m | 10 V/m | d = 0.35 √P 80 MHz to 800 MHz | ||
IEC | 80 MHz to 2.5 GHz |
| d = 0.70 √P 800 MHz to 2.5 GHz | ||
|
|
| |||
|
|
| where P is the maximum output power rating of the | ||
|
|
| transmitter in watts (W) according to the transmitter | ||
|
|
| manufacturer and d is the recommended separation | ||
|
|
| distance in meters (m). |
| |
|
|
| Field strengths from fixed RF transmitters, as | ||
|
|
| determined by an electromagnetic site survey,a | ||
|
|
| should be less than the compliance level in each | ||
|
|
| frequency range.b |
| |
|
|
| Interference may occur in the vicinity of equipment | ||
|
|
| marked with this symbol | . | |
|
|
|
|
|
NOTE 1: Ut is the AC mains voltage prior to application of the test level.
NOTE 2: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 3: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
aField strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the S8 device is used exceeds the applicable RF compliance level above, the S8 device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the S8 device.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.