Activating a Mechanical Emergency Release
Primary Emergency Release
NOTE: For this procedure, refer to
FIGURE 3.5.
NOTE: This procedure will bring the boom down if the hand control is not functioning properly.
To activate the primary emergency release, insert a pen into the hole labeled Emergency Down on the control box.
Secondary Emergency Release
NOTE: For this procedure, refer to
FIGURE 3.6.
NOTE: All patient lift actuators are equipped with a mechanical emergency release. The mechanical release will enable the actuator to retract without power.
NOTE: Use the primary emergency release first before using the secondary emergency release procedure. This procedure should only be used if the primary emergency release procedure is not functioning or is unreachable.
Control Box
Emergency Down
Hole
FIGURE 3.5 Primary Emergency Release
Actuator
Column
Pull Up Red
Tag for
Emergency
Lowering
NOTE: The lift MUST be under a load for the | FIGURE 3.6 Secondary Emergency Release |
| |
mechanical release to function. |
|
To activate the secondary emergency release, pull up on the RED emergency grip and pull down on the boom at the same time.
Reliant™ 440 | 18 | Part No 1141431 |