4. Troubleshooting and Maintenance
Error Display | Cause | Remedy |
|
|
|
| Air plug disconnected. | Insert the plug securely. |
| Refer to section 3.1. | |
|
| |
|
|
|
| Arm cuff not applied correctly. | Apply the arm cuff correctly. |
| Refer to section 3.1. | |
|
| |
|
|
|
| Air is leaking from the arm cuff. | Replace the cuff with the new one. |
| Refer to Chapter 5. | |
|
| |
|
|
|
|
| Repeat measurement. Remain still and do not |
|
| talk during measurement. |
|
| Refer to section 3.3. |
| Cuff over inflated. |
|
| If “E2” appears repeatedly, inflate the cuff | |
|
| manually until it is 30 to 40 mmHg above your |
|
| previous measurement result. |
|
| Refer to section 3.3. |
|
|
|
| The arm cuff was inflated above 299 mmHg | Do not inflate the cuff above 299 mmHg. |
| when inflating the cuff manually. | Refer to section 3.3. |
|
|
|
|
| Repeat measurement. Remain still and do not |
| Movement during measurement. | talk during measurement. |
|
| Refer to section 3.3. |
|
|
|
|
| Remove any clothing interfering with the arm |
| Clothing is interfering with the arm cuff. | cuff. |
|
| Refer to section 3.1. |
|
|
|
| Device error. | Contact your Omron retail outlet or distributor. |
|
|
|
14