| SPECIFICATIONST TLE |
Name: | Omron® IntelliTemp™ Ear Thermometer |
Model: | |
Power source: | DC 3 Volt (lithium battery CR2032) |
Battery life: | Approximately 2500 measurements |
Sensor: | Thermopile |
Temperature indication: In 4 digits, every 0.1°F or °C | |
Accuracy: | 93.2 ~ 96.7°F (34.0 ~ 35.9°C) : ±0.5°F (±0.3°C) |
| 96.8 ~ 102.2°F (36.0 ~ 39.0°C) : ±0.4°F (±0.2°C) |
| 102.3 ~ 108.0°F (39.1 ~ 42.2°C) : ±0.5°F (±0.3°C) |
Measurement range: | 93.2 to 108°F (34 to 42.2°C) |
Weight: | Approx. 1.8 oz (50 g) (including the battery) |
Operating | Ambient temperature: 50 to 104°F (10 to 40°C) |
temperature / humidity: | Relative humidity: 30 to 85% |
Storage | Ambient temperature: |
temperature / humidity: | Relative humidity: 10 to 95% |
Memory: | Memory function displays the previous reading |
*ASTM Standards: | E |
Contents: | Thermometer, Storage Case, Instruction Manual, |
| and 21 Lens Covers |
Accessories: | Omron |
UPC: | 0 73796 33514 4 |
|
|
The specifications may be changed without prior notice.
*ASTM laboratory accuracy requirements for the thermometer only (not including a lens filter) in the display range of 96.8 ˚F to 102.2 ˚F (36 ˚C to 39 ˚C) for infared thermomters is ± 0.4 ˚F (± 0.2 ˚C), whereas for
E
15