Omron M6 Comfort instruction manual Error Messages and Troubleshooting

Page 24

4. Error Messages and Troubleshooting

Display

Cause

Solution

 

 

 

 

Air plug is disconnected.

Insert the air plug securely.

 

Refer to section 3.1.

 

 

 

 

 

 

Arm cuff is applied too loosely.

Apply the arm cuff tighter.

 

Refer to section 3.1.

 

 

 

 

 

 

Air is leaking from the arm cuff.

Replace the arm cuff with a new one.

 

Refer to section 5.3.

 

 

 

 

 

 

 

Repeat measurement. Remain still and do not

 

 

talk during measurement.

 

Movement during measurement and the arm

Refer to section 3.3.

 

 

 

cuff has not been inflated sufficiently.

If “E2” appears repeatedly, inflate the arm cuff

 

 

manually until it is 30 to 40 mmHg above your

 

 

previous measurement result.

 

 

Refer to section 3.3.

 

 

 

 

The arm cuff was inflated above 299 mmHg

Do not inflate the arm cuff above 299 mmHg.

 

when inflating the arm cuff manually.

Refer to section 3.3.

 

 

 

 

 

Repeat measurement. Remain still and do not

 

Movement during measurement.

talk during measurement.

 

 

Refer to section 3.3.

 

 

 

 

 

Remove any clothing interfering with the arm

 

Clothing is interfering with the arm cuff.

cuff.

 

 

Refer to section 3.1.

 

Device error.

Contact your OMRON retail outlet or

 

distributor.

 

 

 

 

 

EN

19

Image 24
Contents Model M6 Comfort Page ABH Check following components Contents Important Safety Information General UsageImportant Safety Information Battery Usage Know Your Device Open the rear cover page to read the followingMonitor Arm cuffDisplay Symbols Irregular Heartbeat SymbolBlood Pressure Colour Indicator Battery Installation PreparationSetting the Date and Time Press the H Date/Time setting buttonPush or button to change the year Press the START/STOP button to turn the monitor offUsing the Device Applying the Arm CuffSecure closed with the fabric fastener How to Sit Correctly Taking a Measurement Select your User ID 1 or Press the START/STOP buttonPress and hold the START/STOP button for more than 3 seconds Using the Memory Function Select your User ID 1 or Press the buttonPress the or button to view the values stored in the memory To View the Average Value To View the Weekly Average Value Morning Weekly Average Page Error Messages and Troubleshooting Error MessagesError Messages and Troubleshooting Troubleshooting Problem Cause SolutionReplace the batteries with new Batteries are empty Ones Maintenance and Storage MaintenanceStorage Do not store the device in the following situationsOptional Medical Accessories Arm cuffPlug the AC adapter into an electrical outlet Specifications Power sourceSpecifications Specifications Warranty Some Useful Information about Blood Pressure Omron Healthcare Europe B.V