4. Error Messages and Troubleshooting
4.2Troubleshooting
Problem | Cause | Solution | |
|
|
| |
| Arm cuff is applied too loosely. | Apply the arm cuff tighter. | |
| Refer to section 3.1. | ||
|
| ||
|
|
| |
| Movement or talking during | Remain still and do not talk during | |
The measurement result is extremely | measurement. | ||
measurement. | |||
high (or low). | Refer to section 3.3. | ||
| |||
|
|
| |
| Clothing is interfering with the arm | Remove any clothing interfering | |
| with the arm cuff. | ||
| cuff. | ||
| Refer to section 3.1. | ||
|
| ||
|
|
| |
| The air connector is not securely | Make sure that the air tube is | |
| connected securely. | ||
| connected into the air jack. | ||
| Refer to section 3.1. | ||
Arm cuff pressure does not rise. |
| ||
|
| ||
| Replace the arm cuff with a new | ||
|
| ||
| Air is leaking from the arm cuff. | one. | |
|
| Refer to section 5.3. | |
|
|
| |
|
| Apply the arm cuff correctly so that | |
Arm cuff deflates too soon. | The arm cuff is loose. | it is firmly wrapped around the arm. | |
|
| Refer to section 3.1. | |
|
|
| |
|
| Inflate the arm cuff so that it is 30 to | |
Cannot measure or the results are | The arm cuff has not been inflated | 40 mmHg above your previous | |
too low or too high. | sufficiently. | measurement result. | |
|
| Refer to section 3.3. | |
|
|
|
20