Troubleshooting

Trouble

Possible cause

 

 

How to correct

 

 

 

 

 

Batteries are worn out.

Replace the batteries with new ones.

Nothing is displayed when

 

 

 

 

(See to Page 6.)

you push the ON/OFF

 

 

 

 

The directions of polarities

 

 

 

button.

Insert the batteries correctly.

 

of inserted batteries

 

(Refer to Page 6.)

 

(+ and -) are wrong.

 

 

 

 

 

 

 

 

 

The wrist cuff is not held

Measure your blood pressure again in

 

correct posture.

 

at the level of the heart.

 

(Refer to Page 9.)

 

 

Measurement cannot be

 

 

 

The wrist cuff is not

Wrap the wrist cuff correctly. (Refer to

made.

wrapped firmly.

Page 7.)

Or the blood pressure

 

 

 

 

Too much stress is applied

Relax yourself and make measure-

values are displayed abnor-

mally high (low).

on your shoulder or arm.

ment.

 

 

 

 

 

 

 

You are talking or moving

 

 

 

 

your hands during meas-

Be still and make measurement.

 

urement.

 

 

 

 

 

 

 

 

The blood pressure values

Refer to the section on "Wrist Blood Pressure Monitor."

differ from the values meas-

(Refer to Page 21.)

 

 

 

ured at brachial.

 

 

 

 

 

 

 

 

 

 

 

 

Blood pressure values are

 

 

 

Contact the store where the

displayed before the monitor

 

 

 

Air is leaking from the wrist cuff.

 

unit was sold or the nearest

inflates to the required

 

 

 

 

OMRON distributor.

value.

 

 

 

 

 

 

 

 

 

 

 

 

Blood pressure value fluc-

Blood pressure value tends to fluctuate depending on the mental

tuates each time of mea-

 

 

 

 

surement.

state during measurement or the measuring time. Breathe deeply to

Or the blood pressure val-

relax yourself and make measurement again.

ues are displayed abnormal-

(Refer to Page 22.)

 

 

 

ly high (low).

 

 

 

 

 

 

 

 

 

 

 

 

If you cannot resolve the problem even after taking the above-mentioned measures, please contact the nearest OMRON dealer or distributor.

Spare parts and accessories

English

 

non-serviceable part because of the necessity of proper calibration after replacement of high-tech

OMRON considers, due to the used high company and quality standards, the main unit as a

 

components.

 

 

The cuff can be replaced by an authorised expert.

 

Specifications

 

 

 

 

 

 

 

 

 

 

Name :

OMRON Fully Automatic Blood Pressure Monitor

 

 

for measurement on the wrist

 

 

 

 

Model :

637IT

 

 

 

 

Display :

Dot matrix LCD display

 

 

 

 

Measurement :

Oscillometric method

 

 

 

 

Measurement Range :

Pressure; 0 to 299 mmHg, Pulse rate; 40 to 180 /min.

 

 

 

 

Accuracy :

Pressure; within +/-3 mmHg, Pulse rate; within +/-5% of reading

 

 

 

 

Inflation :

Automatic inflation with pump

 

 

 

 

Deflation :

Automatic rapid deflation

 

 

 

 

Pressure Detection :

Electrostatic capacity semi-conductor pressure sensor

 

 

 

 

Power supply :

Two LR03 batteries (AAA)

 

 

 

 

Battery Life :

Approximately 300 times of measurement when using alkaline

 

 

batteries (In case of the monitor is used three times a day inflat-

 

 

ing to 170 mmHg at the room temperature of 22°C)

 

 

 

 

Operating Temperature

 

 

and Humidity :

+10° to +40°C, 30 to 85% RH

 

 

 

 

Storage temperature

 

 

and humidity :

-20° to +60°C, 10 to 95% RH

 

 

 

 

Measurable

 

 

circumference of wrist :

135 to 215 mm

 

 

 

 

Weight of Main Unit :

Approximately 150 g (not including batteries)

 

 

 

 

External Dimensions :

78 (W) x 65.5 (H) x 37.2 (D) mm (not including the wrist cuff)

 

 

 

 

Electric Shock Protection :

Internal power supply appliance type B

 

 

 

 

Accessories :

Two LR03 alkaline batteries for monitoring use, storage case,

 

 

Instruction Manual, USB cable, software

 

 

 

 

• In the interest of product improvement, specifications are subject to change without notice.

EU-Representative:

OMRON Healthcare Europe B.V.

=Type B

Kruisweg 577

 

NL-2132-NA Hoofddorp

24

25

Image 13
Contents English Deutsch Nederlands Français Italiano Español Before Measuring Your Blood Pressure Table of ContentsNames and Functions of the Parts Wrist Cuff Battery CoverImportant instructions for obtaining meaningful readings Data Transfer PortHow to Measure Blood Pressure Correctly How to Insert the BatteriesHow to Set the Date and Time How to wrap the cuffCorrect Posture Incorrect Posture How to obtain a correct postureMeasuring your blood pressure and pulse How to Use the Memory Function How to Measure Blood PressureDo not move until the measurement results are DisplayedHow to change the settings How to use the graph functionExplanation of the graph display Systolic blood pressure is higher than 170mmHgLeft or right wrist measurement AlarmAuto memory Reset settings How to use the computer interfaceTime Delete data How to set the time How to create a data file How to use the software Blood pressure at the upper arm and wrist What is blood pressure?Wrist Blood Pressure Monitor Health and blood pressureError indication Maintenance and storageError Cause How to correct Diseases influencing blood pressure valuesSpare parts and accessories TroubleshootingSpecifications Trouble Possible cause How to correct