EN
detect your blood pressure. Before every measurement, the main unit establishes a “zero pressure” equivalent to the air pressure. Then it starts inflating the wrist cuff to 180mmHg or higher until it senses that it has blocked your blood in the artery. After that, the deflation process starts, during which time the main unit detects pressure oscillations generated by
GETTING STARTED
Installing and replacing the batteries
To install the batteries:
1.Slide off the battery cover (Figure 5a).
2.Install the batteries by matching the correct polarity, as shown by Figure 5b. Always use the correct battery type (2 Alkaline LR03
3.Replace the cover (Figure 5c).
Figure 5a | Figure 5b | Figure 5c |
Note: Replace the batteries whenever the weak battery mark ( ) shows, the display is dim, or the display does not illuminate when the power is on. Replace all the batteries at the same time – it is dangerous to mix old and new batteries.
Contact your local waste disposal authority for instructions on how to dispose of used batteries. Used batteries can be harmful to the environment, and should not be thrown out with household trash.
Setting the date, time, and measurement units
It is important to set the clock before using your blood pressure monitor, so that a time stamp can be assigned to each record that is stored in the memory.
To set the date, time, and measurement unit:
1.Press and hold SET for 2 seconds to enter the setting mode.
The setting order is as follows: 12/24 hour format, hour, minute, year month/day or day/month format, month, date, and measurement unit (mmHg or kPa).
Note: Select the measurement unit most applicable for your region. mmHg is used worldwide, including the United States and most of Europe. kPa is primarily used in the People's Republic of China.
2.Press MEM / UP to increase a value or change the setting.
3.Press SET to accept the change and switch to the next setting.
4.To exit the setup menu anytime, press ON.
6