Drive Medical Design 2850-20, 2850-18 manual Orward

Page 5

F ORWARD

Guidance and manufacturer’s declaration - electromagnetic emissions information

The Trident is suitable for use in the electromagnetic environment specified below. The customer or the user of the Trident should assure that it is used in such an electromagnetic environment.

IMMUNITY test IEC 60601 TEST LEVEL

Radiated RF 3 V/rms

IEC 61000-4-6 150 kHz to 80 MHz

Conducted Rf

3 V/m

 

IEC 61000-4-3

80 kHz t

o 25 MHz

 

 

 

Compliance level

Not applicable

10 V/m

Electromagnetic environment - guidance

Portable and mobile RF communications equipment should be used no closer to and part of the (ME EQUIPMENT or ME SYSTEM), including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.

Recommended separation distance

d = [ 3,5 / V1 ] √P

d = [ 3,5 / E1 ] √P 80 MHz to 800 MHz

d = [ 7 / E1 ] √P 800 MHz to 2.5 GHz

where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m).

Field strengths frrom fixed RF transmitters, as deter-mined by an electromagnetic site survey, ª should be less than

the compliance level in each frequency range. B

Interference may occur in the vicinity of equipment marked with the following symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

aField strengths from fixed transmitters, such as base stations for radio (cellular / cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electrom agnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the (ME EQUIPMENT or ME SYSTEM) is used exceeds the applicable RF compliance level above, the (ME EQUIPMENT or ME SYSTEM) should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the (ME EQUIPMENT or ME SYSTEM).

bOver the frequency range 150 kHz to 80 MHz, field strengths should be lsee than [V1] V/m.

5

Image 5
Contents Trident Power Wheelchair Contents Cispr ForwardForward Orward Product Direction ForwardSpecifications SpecificationsDrive Page Unpacking & Assembly Unpacking & Assembly Unpacking & Assem BLY Unpacking & Assembly ADJ Ustment Adjustment1Fig.15-2 Adjustment Drivepush Operati onOperati on Operati on Operati on Changing the Battery Operati on Operati on Wiring diagra m Control system cable connections Control system fault Operation Operation Transport Ation & Storage MAI Ntenance Aintenance Check Warranty Conditions Warranty