Problem/Possible cause

Solution

Displays error message: Check ResMed 30/90W Power Supply Unit and fully insert the connector

The DC plug is partially inserted into the back of the device or inserted too slowly.

A non-ResMed power supply unit is connected to the device.

The power supply unit is being covered by bedding.

Fully insert the DC plug.

Remove the power supply unit and replace with a ResMed power supply unit.

Make sure that the power supply unit is free from bedding, clothes or other objects that could cover it.

Displays error message: Tube blocked, please check your tube

Air tubing is blocked.Check your air tubing and remove any blockages. Disconnect the power cord and then reconnect it to restart the device.

Displays error message: High leak, please check system setup and all connections

There is excessive leak.

Adjust position of mask and headgear.

 

Air tubing is not connected properly. Connect firmly at both

 

ends.

The following message is displayed on the LCD after you try to update settings or copy data to the SD card: Card error, please remove SD card and contact service provider

SD card is not inserted correctly.

Ensure that the SD card is inserted correctly.

You may have removed the SD card

Reinsert the SD card and wait for the home screen or the

before settings were copied to the

“Settings updated successfully, press any key”

device.

message to appear on the LCD.

 

Note: This message only appears once. If you re-insert the

 

SD card after you have updated your settings, the message

 

will not be re-displayed.

The following message is NOT displayed on the LCD after you try to update the settings using the SD card: Settings updated successfully, press any key

The settings were not updated.

Contact your clinician/service provider immediately.

6 English

Image 8
Contents Vpap ST Page S9 Vpap ST indications for use S9 Vpap S indications for useS9 AutoSet CS indications for use S9 Auto 25 indications for useNavigating the menus Masks and humidifiers SetupControl panel Cleaning and maintenance Getting startedTroubleshooting Problem/Possible causeProblem/Possible cause Solution Will not be re-displayed Technical specifications SD cardDevice can be higher than the temperature of the room Symbols Pneumatic flow pathEnvironmental information Limited warrantyServicing English English English
Related manuals
Manual 2 pages 23.72 Kb

AutoSet CS, S9 Auto 25, VPAP S specifications

ResMed, a leader in sleep and respiratory care, has developed a range of advanced devices designed to enhance the quality of life for those suffering from sleep apnea and other respiratory conditions. Among these, the ResMed AutoSet CS, S9 Auto 25, and VPAP S stand out due to their innovative features and technologies aimed at providing personalized therapy.

The ResMed AutoSet CS is engineered to offer adaptive therapy for patients facing complex sleep apnea. This device utilizes advanced algorithms that continuously assess the patient’s breathing patterns throughout the night. By analyzing flow limitation, apnea, and hypopnea events in real time, it adjusts the pressure delivered to ensure optimal airflow. The AutoSet CS is particularly beneficial for those who experience varying levels of airway obstruction, providing a tailored response that improves overall comfort and efficacy.

The S9 Auto 25 is another remarkable device, distinguished by its user-friendly interface and sleek design. This auto-titrating CPAP device automatically adjusts pressure levels based on the patient’s needs throughout the night. It incorporates ResMed’s QuietAir technology, ensuring a quieter operation, which promotes a peaceful sleep environment. Additionally, the S9 Auto 25 features an integrated humidification system that delivers moisture to the airflow, preventing dryness and irritation, which are common complaints among CPAP users.

The VPAP S represents ResMed’s commitment to providing bi-level positive airway pressure therapy. It offers two distinct pressure settings: one for inhalation and a lower setting for exhalation. This dual-pressure mechanism allows for a more natural breathing experience, particularly beneficial for those with more severe forms of apnea. The VPAP S also features Easy-Breathe technology, which smooths out pressure transitions, making it easier for patients to breathe naturally and comfortably through the night.

Key characteristics shared by these devices include advanced data tracking and reporting capabilities. Users can access valuable insights into their therapy through user-friendly interfaces or companion apps, facilitating better management of their condition. Each model is designed with portability in mind, ensuring patients can receive effective therapy whether at home or on the go. With ResMed’s commitment to innovation, the AutoSet CS, S9 Auto 25, and VPAP S are pivotal in transforming the landscape of sleep apnea treatment, providing patients with effective, comfortable, and adaptable therapy solutions.