Swift FX Bella Gray Swift FX
Problem / possible cause | Solution |
|
|
Ears are sore |
|
|
|
Swift FX Bella Gray headgear is not fitted correctly.
Readjust the straps to ensure they are not overtightened or twisted. If problem persists talk to your clinician.
Swift FX Bella Gray headgear is not suited for your ear type.
Replace the Swift FX Bella Gray headgear with the supplied Swift FX headgear.
Technical specifications
curverebreathing. As a result of manufacturing variations, the vent flow rate may vary.
| 60 |
|
|
|
|
|
|
| Pressure | Flow |
|
|
|
|
|
|
|
| (cm H2O) | (L/min) | |
(L/min) | 50 |
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
| ||
40 |
|
|
|
|
|
|
| 4 | 20 | |
|
|
|
|
|
|
|
|
|
| |
rate | 30 |
|
|
|
|
|
|
| 8 | 29 |
|
|
|
|
|
|
|
| |||
flow | 20 |
|
|
|
|
|
|
| 12 | 37 |
10 |
|
|
|
|
|
|
| |||
Vent |
|
|
|
|
|
|
|
|
| |
0 | 6 | 8 | 10 | 12 | 14 | 16 | 18 | 16 | 43 | |
| 4 | 20 |
| |||||||
|
|
|
| Mask pressure (cm H2O) |
|
| 20 | 49 | ||
|
|
|
|
|
|
|
|
Dead space | Physical dead space is the empty volume of the mask |
information | to the end of the swivel. Using the large cushions it is |
| 106 mL. |
|
|
Therapy pressure | 4 to 20 cm H2O |
Resistance | Drop in pressure measured (nominal) |
| at 50 L/min: 0.4 cm H2O |
| at 100 L/min: 1.4 cm H2O |
Environmental Operating temperature: +5°C to +40°C
conditions Operating humidity: 15% to 95% relative humidity
Storage and transport temperature:
Storage and transport humidity: up to 95% relative humidity
10