Reprocessing the mask between patients
This mask should be reprocessed when used between patients. Cleaning, disinfection and sterilisation instructions are available from the ResMed website, www.resmed.com/masks/sterilization.If you do not have internet access, please contact your ResMed representative.
Troubleshooting
Problem / possible cause | Solution |
Pillows won’t seal properly, are uncomfortable or cause red marks
English
Pillows may have been fitted incorrectly, adjusted incorrectly or the headgear is too tight.
Check that the ResMed logo on top of the pillows is facing outwards. Carefully follow the fitting instructions. Readjust the headgear to ensure it is not over tightened.
Inner wall position of one or both | Squeeze pillow sides to release |
pillows is incorrect. | inner wall. The correct position |
| is shown in the reassembly |
| instructions. |
Pillows are the wrong size. Pillows may be dirty.
Mask is too noisy
Talk to your clinician.
Clean pillows according to the instructions.
System is assembled incorrectly. | Disassemble the mask, then |
| reassemble according to the |
| instructions. |
System is leaking air.
Vent is blocked or partially blocked.
Refit or reposition you mask. Adjust your headgear.
Check your mask is assembled correctly.
If the vent requires cleaning, use a soft bristle brush.