Omega TX91 manual Calibration Values for the TX92

Page 28

4Calibration Instructions

Table 4-1. Calibration Values for the TX92

 

Temperature

 

mV INPUT

 

mV INPUT

 

Input Range

Model

REF 32°F

Model

REF 32°F

 

Zero/Span

TX92

Zero/Span

TX92

Zero/Span

 

0/1200°F

-J2

-0.855/14.906

-K2

-0.692/13.819

 

0/1300°F

-J3

-0.885/17.947

-K3

-0.692/16.092

 

0/1500°F

-J4

-0.885/14.108

-K4

-0.692/10.560

 

0/1750°F

-J5

-0.885/20.406

-K5

-0.692/15.295

 

0/1000°F

-J6

-0.885/29.515

-K6

-0.692/22.251

 

0/1200°F

-T2

-0.674/13.967

-E2

-1.026/15.869

 

0/1300°F

-T3

-0.674/16.647

-E3

-1.026/19.708

 

0/1500°F

-T4

-0.674/12.572

-E4

-1.026/17.942

23

0/1750°F

--

--0.674/12.572

-E5

-1.026/26.858

0/1000°F

--

--0.674/12.572

-E6

-1.026/40.056

Image 28
Contents User’sGuide Servicing North America Return Requests / Inquiries Introduction Calibration Instructions Introduction General DescriptionIntroduction TX91 Thermocouple Transmitter Introduction Introduction Straight line Approximation of Curve Features Models AvailableTX91 Models Available Model Number Description Unpacking Installation Mounting the TX91Assembly of the TX91 Transmitter Inside RT Mounting Track in inches TX90-BR Mounting Bracket in inches Mounting Track RT Bracket TX90-BR Wiring the TX91 Refer to Figure Wiring Diagram for the TX91 Potentiometers Calibration Instructions Equipment RequiredSet-up of Equipment Calibration Procedures Refer to Figure Calibration Instructions Calibration Set-Up Calibration Set-Up. Alternate Calibration Values for the TX92 Troubleshooting Guide Current rating on the power supplyTroubleshooting Guide Accessories Model No DescriptionSpecifications GeneralInput Intrinsically Safe Interconnection Diagram WARRANTY/DISCLAIMER Shop online at omega.com