Omega CL24, CL25, CL27, CL23A, CL26 manual Opr, Meter Mode Operation Thermocouples Set-Up

Page 15

METER: (unit functioning as a thermometer)

K:(K-type thermocouple)

°F:

(Fahrenheit readings)

OPR:

(Instrument is operational)

If no input thermocouple is connected to the unit at this time, the numeric segments will indicate "OPEN".

Plugging a thermocouple into the meter jack (connector T2, on the top right side) will give actual thermocouple temperature readings.

NOTE: To obtain full accuracy, allow 1-2 minutes after connect- ing a thermocouple plug, for thermal setting.

If the instrument is to be used as a thermometer or ohmmeter, refer to the Meter Mode Operation section for detailed instructions. For use as a tem- perature-calibrator or ohms simulator, refer to Calibrator Mode Operation.

3.METER MODE OPERATION (Thermocouples) Set-Up:

1.Turn on instrument by depressing the ON/OFF key.

2.Check that the BAT annunciator turns off following the power-up LCD test. If it does not turn off, less than 10% of battery remains. Refer to Battery Installation/Replacement instructions.

3.If the meter is already in the METER mode, the METER annunciator will be on. Otherwise the calibrator mode is enabled and the CALIB annunci- ator is on. Depress the METER key to return to meter mode.

4.Connect an input thermocouple (TC) to connector T2. Select the appro- priate TC Type with the select-key. A display annunciator will indicate the selected TC-type.

NOTE: Space and legibility constraints on the display preclude the use of dedicated annunciators for the less common tungsten- rhenium thermocouple types (G, C & D). Instead, these types are annunciated momentarily on the alpha-numeric display when first selected. Both T and R annunciators remain on to indicate a Tungsten-Rhenium selection. To recall the actual G, C, D selec- tion, turn the unit off, then on again for another momentary read- out.

5.Select temperature scale with °F/°C key. The corresponding annunciator is enabled.

13

Image 15 Contents
User’sGuide Omega.comTM Table of Contents General Information Common SpecificationsTemperature Coefficient Page Model CL25/CL25EC Unique SpecificationsModel CL23A/CL23EC Model CL24Features Manual AddendaModel CL27/CL27EC Safety Information Safety PrecautionsUnpacking Your Unit Page Battery Installation Battery INSTALLATION/REPLACEMENTOperation with Rechargeable Battery Memory BackupInitial Turn on Diagnostics and Error MessagesProblem Likely Cause Operating InstructionsKeypad OPR Meter Mode Operation Thermocouples Set-UpHold Meter Mode Operation RTD, Thermistor & Ohms Set-UpData Logging and Recall To Recall Readings To Save ReadingsCalibrator Mode Operation Thermocouples Set-Up To Erase Data Locations Meter ModeCHANGE/ENTER, 2, 1, 2, 0, CHANGE/ENTER Calibrator Mode Operation RTD, Thermistor & Ohms Set-UpWire Resistance Simulation To Store Outputs Storage and Recall of Calibrator SettingsTo Recall Outputs To Erase Set-Up Locations Calibrator ModeRamp Function CL25 and CL27 Step FunctionAmbient Conditions Service InformationCalibration Procedures Thermocouple Calibration Test Equipment RequiredPage Calibration Adjustments Output Adjust Reading Meter-Mode Zero & Gain Adjustments Sensor StepInput Adjust Reading Calibrator-Mode Zero & Gain Adjustments SensorCalibration Setup, Copper Mode Test Equipment Required Resistance CalibrationResistance Calibration Setup RTD-100 390.26Ω RTD-Thermistor Sensor Unit Step Output Input Adjust Verify ReadingInput Adjust Verify Reading Calibration of Meter-Mode Resistance FunctionsCalibration Verification Equipment Required Model Software RevisionPage Page Alloy Mode Verification Procedure for Thermocouple Functions Page Mechanical Parts Diagram Calibration Procedures Page Page Page USAWARRANTY/DISCLAIMER Temperature