Omega CL24, CL25, CL27, CL23A, CL26 Service Information, Calibration Procedures, Ambient Conditions

Page 25

SERVICE INFORMATION

WARNING

All service information is intended for qualified electronic mainte- nance personnel only.

1. CALIBRATION PROCEDURES

This is a two-part procedure. Products with thermocouple functions only, fol- low the Thermocouple Calibration procedure below. Products with resistance functions (RTD, thermistor, ohms) follow first the Thermocouple Calibration procedure, then the Resistance Calibration procedure.

A.THERMOCOUPLE CALIBRATION Test Equipment Required:

1.Thermocouple simulator (OMEGA CL521, or equivalent) calibrated to ITS-90.

2.Microvolt DMM (OMEGA OM7551, or equivalent)

3.Calibration cables, per Figures 11, 12, 13.

NOTE: The copper constantan cables used in Figures 12 and 13 need to be calibrated. Voltage errors in these cables should be com- pensated at the simulator.

4. Calibration cover (OMEGA 820-307-4).

Ambient Conditions:

Units should be calibrated at an ambient temperature of 23°C ±1°C, with rel- ative humidity less than 80%.

Preparation for Calibration:

11. Remove battery from unit under test (U.U.T.). Refer to Battery Installation/Replacement section in the manual.

12. Remove bottom-cover from U.U.T. Refer to Disassembly Instructions.

13. Remove calibration-jumpers J1 and J2 from printed-circuit board.

NOTE: Removal of J1, J2 causes partial loss of previously stored calibration data. Do not remove these jumpers unless recalibra- tion is intended.

23

Image 25 Contents
User’sGuide Omega.comTM Table of Contents General Information Common SpecificationsTemperature Coefficient Page Model CL23A/CL23EC Unique SpecificationsModel CL24 Model CL25/CL25ECManual Addenda FeaturesModel CL27/CL27EC Safety Precautions Safety InformationUnpacking Your Unit Page Battery Installation Battery INSTALLATION/REPLACEMENTOperation with Rechargeable Battery Memory BackupProblem Likely Cause Diagnostics and Error MessagesOperating Instructions Initial Turn onKeypad OPR Meter Mode Operation Thermocouples Set-UpHold Meter Mode Operation RTD, Thermistor & Ohms Set-UpData Logging and Recall To Recall Readings To Save ReadingsCalibrator Mode Operation Thermocouples Set-Up To Erase Data Locations Meter ModeCHANGE/ENTER, 2, 1, 2, 0, CHANGE/ENTER Calibrator Mode Operation RTD, Thermistor & Ohms Set-UpWire Resistance Simulation To Store Outputs Storage and Recall of Calibrator SettingsTo Recall Outputs To Erase Set-Up Locations Calibrator ModeRamp Function CL25 and CL27 Step FunctionCalibration Procedures Service InformationThermocouple Calibration Test Equipment Required Ambient ConditionsPage Calibration Adjustments Input Adjust Reading Meter-Mode Zero & Gain Adjustments Sensor StepCalibrator-Mode Zero & Gain Adjustments Sensor Output Adjust ReadingCalibration Setup, Copper Mode Test Equipment Required Resistance CalibrationResistance Calibration Setup RTD-100 390.26Ω Input Adjust Verify Reading Output Input Adjust Verify ReadingCalibration of Meter-Mode Resistance Functions RTD-Thermistor Sensor Unit StepCalibration Verification Equipment Required Model Software RevisionPage Page Alloy Mode Verification Procedure for Thermocouple Functions Page Mechanical Parts Diagram Calibration Procedures Page Page Page USAWARRANTY/DISCLAIMER Temperature