2. SET-UP PROCEDURE

2.1Battery Installation

All MAXO2 analyzer units are powered by two, AA alkaline batteries (2 x

1.5Volts) and are shipped without the batteries installed. The battery compartment is accessible from the back side of the unit. To install the batteries:

1)With the thumb, press down on the center of the battery compartment cover and slide the cover off of the instrument case.

2)Install the two, AA, alkaline batteries (2 x 1.5 Volts) in the unit, observing the orientation shown on the label inside the compartment.

3)Slide the battery compartment cover back onto the case. Make sure the tabs on the cover snap into position, securing the cover flush against the case.

When batteries are installed in the MAXO2 analyzer, the unit initiates a self-diagnostic test. All segments of the LCD readout are turned on for approximately 2 seconds. When the diagnostic test is completed successfully, the word "OFF" will appear on the display, indicating that the unit is ready to be calibrated.

2.2Calibrating the MAXO2 Analyzer 2.2.1 Before You Begin

The MAXO2 Oxygen Analyzer should be calibrated upon receipt. Thereafter, Maxtec recommends calibration on a weekly basis. However, more frequent calibration will not adversely affect product performance. More frequent calibration is required when:

The temperature of the gas stream changes by more than 3 degrees Celsius.

Changes in elevation result in calibration error of approximately 1% of reading per 250 feet. In general calibration of the instrument should be performed when the geographic elevation at which the product is being used changes by more than 500 feet.

In addition, calibration is recommended if the user is unsure when the last calibration procedure was performed or if the measurement value displayed is in question.

5

Image 11
Contents Model OM-25A Page Thank YOU PrefacePage Iii Spare Parts and Accessories Warranty Table of Contents System OverviewSET-UP Procedure Base Unit Specifications Sensor SpecificationsBase Unit Description CD Readout Components Description please refer toLCD Readout a Keypad Sample Inlet Connection OM-25ASample Inlet Connection Calibrating the MAXO2 Analyzer 2.2.1 Before You Begin Battery InstallationTo Calibrate the MAXO2 Analyzer Automatic Calibration to Room Air Pressure Effect Effects of TemperatureOperation Instructions Humidity EffectPage Problem Solving Base Unit Specifications Calibration Techniques in Pressurized Systems Exposure to Anesthetic GasesCalibration Errors Accessories Spare Parts and AccessoriesWarranty Page Email sales@maxtecinc.com