Hoya OM-25A manual Effects of Temperature, Pressure Effect

Page 14

4) Release both the CALIBRATE key and the key.

The unit will automatically enter the “LOCKED” condition and return to normal operation.

2.2.4 Factors Influencing Calibration

The primary factors influencing the MAXO2 analyzer are temperature, pressure, and humidity.

Effects of Temperature

The MAXO2 analyzer will hold calibration and read correctly within ±3% when in thermal equilibrium within the operating temperature range. The device must be thermally stable when calibrated and allowed to thermally stabilize after experiencing temperature changes before readings are accurate. For these reasons, the following is recommended:

1)Allow adequate time for the sensor to equilibrate to a new ambient temperature.

2)For best results, perform the calibration procedure at a temperature close to the temperature where analysis will occur.

Pressure Effect

Readings from the MAXO2 analyzer are proportional to the partial pressure of oxygen. The partial pressure of Oxygen (PO2) is equal to the percentage of oxygen (%O2) times the absolute pressure (AP) at which the sample enviroment is measured. (PO2=%O2 x AP). Thus the readings are proportional to the concentration if the pressure is held constant. Flow rate of sample gas can affect pressure at the sensor in that back pressure at the sensing point may change. For these reasons, the following is recommended:

1)Calibrate the MAXO2 analyzer at the same pressure as the sample gas.

2)If sample gases flow through tubing, use the same apparatus and flow rates when calibrating as when measuring.

3)The MAXO2 analyzer oxygen sensor has been validated at pressures up to 2 atmospheres absolute. Calibration or operation above this pressure is beyond the intended use.

8

Image 14
Contents Model OM-25A Page Preface Thank YOUPage Iii Base Unit Specifications Sensor Specifications Table of Contents System OverviewSET-UP Procedure Spare Parts and Accessories WarrantyBase Unit Description Components Description please refer to CD ReadoutOM-25A LCD Readout a Keypad Sample Inlet ConnectionSample Inlet Connection Battery Installation Calibrating the MAXO2 Analyzer 2.2.1 Before You BeginTo Calibrate the MAXO2 Analyzer Automatic Calibration to Room Air Effects of Temperature Pressure EffectHumidity Effect Operation InstructionsPage Problem Solving Base Unit Specifications Exposure to Anesthetic Gases Calibration Techniques in Pressurized SystemsCalibration Errors Spare Parts and Accessories AccessoriesWarranty Page Email sales@maxtecinc.com