Hoya OM-25A manual Automatic Calibration to Room Air

Page 13

8)When the calibration value is set, press the CALIBRATE or LOCK/UNLOCK key again to accept the calibration setting and return to normal operation.

Note: If the message "CAL", followed by the message "Er" flashes on the display after entering the desired calibration value, the system has determined that the entered value will not allow operation within the specified output range of the sensor. This situation may occur if:

a)the operator has inadvertently entered the wrong concentration for the calibration gas.

b)the concentration of the calibration gas is not correct.

c)the sensor is in need of replacement.

d)the operator attempted to adjust the analyzer before allowing sufficient time for the calibration gas to purge out the previous sample.

e)the flow and pressure of the calibration gas was not properly regulated.

Check these items and repeat calibration. If calibration error continues to occur, contact the service department of the distributor from which the unit was purchased, or you may call Maxtec's Customer Service Department directly.

2.2.3 Automatic Calibration to Room Air

The MAXO2 analyzer can quickly be calibrated to room air (20.9%) using a quick-key shortcut command. This function saves time by setting the calibration value to 20.9% without scrolling the display. To use this function:

1)Introduce room air to the sensor at a rate of 1-5 liters per minute (2 liters per minute is recommended) and allow the reading to stabilize.

2)Press the LOCK/UNLOCK key to unlock the keypad.

3)Press and hold down the CALIBRATE key. When the "%" sign starts to flash, press the arrow key to set the calibration value to 20.9%.

7

Image 13
Contents Model OM-25A Page Thank YOU PrefacePage Iii SET-UP Procedure Table of Contents System OverviewBase Unit Specifications Sensor Specifications Spare Parts and Accessories WarrantyBase Unit Description CD Readout Components Description please refer toLCD Readout a Keypad Sample Inlet Connection OM-25A Sample Inlet Connection Calibrating the MAXO2 Analyzer 2.2.1 Before You Begin Battery InstallationTo Calibrate the MAXO2 Analyzer Automatic Calibration to Room Air Pressure Effect Effects of TemperatureOperation Instructions Humidity EffectPage Problem Solving Base Unit Specifications Calibration Techniques in Pressurized Systems Exposure to Anesthetic GasesCalibration Errors Accessories Spare Parts and AccessoriesWarranty Page Email sales@maxtecinc.com