•When power is turned ON, all display marks appear for approximately 1 second.
•When the "o" is displayed the unit is ready for measurement ltthk”V” mark appears, release air in the arm cuff by squeezing it or removing the hose plug.
Approx. 1 second
2.Attach the arm cuff to the left arm for best results.
3.Pressurize the arm cuff with the inflation bulb.
• Pressurize the arm cuff to approximately
pressure to be measured.
1pl
II
4.When Pressurization is Completed, the automatic exhaust mechanism will gradually reduce arm cuff pressure, and the “v “ mark will appear indicating that
measurement is in progress.
1) Pressure value appears on the !eft, and exhaust speed on the right side of the display.
*Remain still during measurement.
2) When a pulse is detected, the pressure value moves to the right side of the display. At this time the “ v” mark flashes at the same rate as pulse rate, and the buzzer sounds when the pulsation is sufficient.
F l a s h i n g]~~fi:~:n’+ v | Pressure |
3)When the measurement is completed and buzzer sounds (beep), Systolic pressure is displayed on the left and Diastolic pressure on the right side of the display.
*When a measurement is made with insufficient pressure the “Err” mark appears. Pressurize to a valve
5. Push the air release valve to completely release the air from within the cuff.
6. Turn Power OFF
Press the Power switch to turn power OFF. (See Diagram A)
7. Previous Measurement Display
When the Memory switch is pressed, the previous measurement value is shown for about 5 seconds.
8. Automatic Power OFF Function
If this unit is left on after measurement an Automatic