Care and maintenance

●● Do not apply excessive force to the instrument or drop the instrument. - Doing so may cause damage.

●● Measurements may be affected if the device is used close to televisions, microwave oven, X-ray equipment or other devices with strong electrical fields. Measure blood pressure at a sufficient distance from such devices or switch them off.

●● If the instrument is stored at temperatures below the freezing point, do not use it immediately. Leave it in a warm place for at least an hour before use.

-If the instrument is not allowed to warm up, it may not pressurize. ●● Do not insert dust or foreign objects into the instrument.

-Doing so may result in damage.

●● If the instrument becomes dirty, clean it with a soft cloth moistened with warm water or soapy water. Do not use alcohol, benzene, or paint thinner.

- Use of such chemicals may result in cracking or discoloration.

●● For storage, avoid high temperatures, high humidity, and direct sunlight. - Storage under such conditions may result in damage.

●● The cuff cannot be washed.

Specifications

Power source

DC 6 V (4 AA/LR6 size alkaline batteries)

 

AC 120 V 60 Hz (when using AC adapter)

Display

Digital LCD

 

 

Method of measurement

Oscillometric system

Measurement range

Pressure: 0 mmHg to 280 mmHg

 

Pulse rate: 30 to 160 beats per minute

Accuracy

Pressure: Within ±3 mmHg

 

Pulse rate: Within ±5 %

Operation temperature/humidity range

50 °F to 104 °F (10 °C to 40 °C), 30 % to 85 % RH

Storage temperature/humidity range

–4 °F to 140 °F (–20 °C to 60 °C), 10 % to 95 % RH

Measurable arm circumference

Approx. 7-3/4" to 15" (20 cm to 38 cm)

 

 

Weight

Approx. 10.9 oz. (310 g) (not including batteries)

Dimensions

3" × 5" × 4-1/2" (7.9 cm × 13.0 cm × 11.5 cm)

 

 

Protection against

Internally powered equipment, Type BF applied part

electric shock

 

 

 

The blood pressure monitor might not perform to specifications if stored or used outside the specified temperature and humidity ranges.

AC adapter

Input

AC 120 V 60 Hz

Output

DC 6 V 500 mA

 

 

21

Image 22 Contents
Upper Arm Blood Pressure Monitor Instrucciones de UsoBefore USE Table of contents Basics of Blood Pressure IntroductionImportant Instructions Before Use Before USE Precautions to Ensure Safe, Reliable Operation Systolic Blood Pressure SBP Parts identification Main unitCuff Systolic blood Average mark Battery icon Record mark PulseTrend graph Pulse indicator PressureReplace batteries when When using the main unit with the AC adapterInserting batteries Be sure to use only the exclusive AC adapter EW-2B02Adjust the year Setting the time and dateHold for 3 seconds or longer Fit the cuff to the upper arm Points when wrappingWrapping the cuff Wrap the cuffGetting into the measurement position MeasurementsTaking a reading Measuring/Recording measurements StartAbout each indicators AdviceReading the blood pressure display Deleting all data stored in memory Recalling readings from memory To complete the measurement Press Weekly average All, AM, PM Comparing with old measurementsMonthly average All Confirm the previous dataTrend Graph Upper Systolic blood pressure How to read the graphsAll data All, AM, PM Trend Graph Lower Diastolic blood pressureUnplug the plug from the main unit Place the main unit and the cuff into the accessory pouchStorage Close the hook and loop fastenerCare and maintenance SpecificationsExplanation of rating plate symbols First Troubleshooting Indicator does not light when an arm is movedPresence FCC Statement Panasonic Corporation of North America