Additional Configuration Options and Features
The TX787 and T788 is capable of performing many functions in addition to the simple example outlined in the Quick Start exercise.
Dynamic Input Filtering
Filtering is provided to optimize the stability and response time of the transmitter output signal. The filter is set by specifying a Filter Band and a Damping Time Constant. To simplify the definition of the filter values, click on the “Set to Optimal” button on the configuration screen.
Filter Band
The Filter Band allows the transmitter to be configured to react quickly to significant changes in input while smoothing out small changes. This band defines the range above and below the current input reading within which the Damping Time Constant will be applied. It is defined as a percent of the configured span of the transmitter. For example, a 0.5% Filter Band on a transmitter configured for an input range of 200 to 700 deg F would mean that the Damping Time Constant will be applied to input changes of less than 2.5 deg but that changes of more than 2.5 deg would be immediately reflected in the output. In general, the larger the full scale range, the smaller the Filter Band should be. Conversely, a larger Filter Band is generally appropriate for smaller input ranges. Clicking the “Set to Optimal” button triggers the calculation of an appropriate setting based on the currently defined input range.
Damping
Damping defines the length of time, in seconds, that the input will be averaged over to determine the output value when the input change is less than the Filter Band. Changes in the input that are greater than the Filter Band bypass the Damping filter and are immediately reflected in the output. The default value of 15 seconds, when used with the optimal Filter Band yields a very stable output while maintaining a fast response to significant changes in the input.
Input/Output Trimming
Because of slight variations between sensing elements or other loop characteristics, it is sometimes beneficial to “trim” the transmitter so that its output corrects for these effects. If the measurement error is known, the transmitter may be trimmed
Dynamic Sensor Simulation
This option allows a loop to be tested dynamically verifying control logic as well as wiring. The transmitter’s current loop output can be controlled via four parameters: simulate an input value in raw units (ohm, mV, etc.); simulate an input value in engineering units (e.g. 450 DEG); set an ouput in percent; or set an ouput in mA. After an initial simulation is selected, the ouput may be incremented or decremented in 1% or 10% steps.
Other Transfer Functions
The transfer function is the “formula” that the transmitter uses to determine the appropriate output signal for a given input signal. A “Linear” function means that output is changed in a straight line fashion as the input changes within the configured input zero and full scale range: 0% of range input yields 4 mA output, 25% of range input yields 8 mA, etc. A “Standard” function means that the transmitter will apply a standard linearization curve for temperature inputs.
RTD Sensor Matching
The accuracy of RTD temperature sensing can be increased significantly when the temperature transmitter is “matched” to the individual sensor using a method called Callender Van Dusen linearization. Doing so requires entering the sensor’s Alpha, Beta (measured for temperatures above zero and for temperatures below zero), Sigma, and Ro values into the transmitters' configuration. These values can be requested from the manufacturer when the sensor is purchased or may be measured by a testing lab.
Square Root
Select this function if the transmitter’s output is to vary with the square root of the input.
Custom Polynomial
The transmitter can use up to a 10th order polynomial equation (y = A0 + A1x + A2x2 + … + A10x10) as the transfer function by inputting the coefficients into the transmitter’s configuration.
Custom Table
The transmitter can use a User Defined table by specifying a .tbl file. The data in the file should be two columns, tab delimited. Up to 140 data pairs may be listed. The input values in column one must be equally spaced.
Defining The Transmitter’s User Fields
Each unit may be assigned an I.D. Tag name of up to eight characters, Job Number up to eight characters, and a “Message” of up to sixteen characters; stored in the transmitter’s memory. Click on the desired field, enter the text, and click the download icon.
| OMEGAnetSM |
| Internet |
| |
|
| http://www.omega.com |
| info@omega.com |
|
|
|
|
|
|
|
USA: | One Omega Drive, Box 4047 |
|
| ||
ISO 9001 Certified | Stamford, CT |
|
| ||
|
| Telephone: (203) | FAX: (203) | ||
|
|
|
| ||
Canada: | 976 Bergar |
|
| ||
|
| Laval (Quebec) H7L 5A1 |
|
| |
|
| Telephone: (514) | FAX: (514) | ||
|
|
|
| ||
USA and Canada: | Sales Service: | ||||
|
| Customer Service: | |||
|
| Engineering Services: | |||
|
| TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA | |||
Mexico and | Tel: (95) | FAX: (95) | |||
Latin America: | En Espanol: (203) | ||||
Benelux: | Postbus 8034, 1180 LA Amstelveen, The Netherlands | ||||
|
| Tel: (31) 20 6418405 | FAX: (31) 20 6434643 | ||
|
| Toll Free in Benelux: 06 0993344 |
|
| |
|
|
|
| ||
Czech Republic: | Ostravska 767, 733 01 Karvina |
|
| ||
|
| Tel: 420 (69) 6311627 | FAX: 420 (69) 6311114 | ||
|
|
|
| ||
France: | 9, rue Denis Papin, 78190 Trappes |
|
| ||
|
| Tel: (33) | FAX: (33) | ||
|
| Toll Free in France: |
|
| |
|
|
|
| ||
Germany/Austria: | Daimlerstrasse 26, |
|
| ||
|
| Deckenpfronn, Germany |
|
| |
|
| Tel: 49 (07056) 3017 | FAX: 49 (07056) 8540 | ||
|
| Toll Free in Germany: 0130 11 21 66 |
|
| |
|
|
|
| ||
United Kingdom: | 25 Swannington Road, | P.O. Box 7, Omega Drive, | |||
IS0 9002 Cerfified | Broughton Astley, Leicestershire, | Irlam, Manchester, | |||
|
| LE9 6TU, England | M44 5EX, England | ||
|
| Tel: 44 (1455) 285520 | Tel: 44 (161) | ||
|
| FAX: 44 (1455) 283912 | FAX: 44 (161) |
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice.
WARNING: These product are not designed for use in, and should not be used for, patient connected applications.