TROUBLESHOOTING
Warning: If you experience a problem with your lift and are unable to service it yourself, contact your
Symptom | Possible Cause | Action |
|
|
|
| Batteries discharged | Charge batteries |
Control box is not responding | Battery charger defective | Replace battery charger |
and relay is not clicking | Batteries defective | Replace batteries |
| Hand control defective | Replace hand control |
|
|
|
| Actuator plug not properly | Push actuator plug firmly into |
Actuator does not run, | seated | control box |
but relay is clicking | Actuator defective | Replace actuator |
| Control box defective | Replace control box |
|
|
|
Actuator goes up or down, but | Hand control defective | Replace hand control |
not both | Control box defective | Replace control box |
|
|
|
Cannot lift patient up and |
|
|
battery indicator shows no | Batteries discharged | Charge the batteries |
GREEN light |
|
|
|
|
|
|
| Check the connection between the |
Lift does not work and the battery |
| actuator and the control box |
indicator shows full charge |
| Check connection between Pendant |
|
| & control box |
|
|
|
|
| Make sure the emergency stop |
Lift does not work and the |
| button is released |
battery indicator shows no light |
| Check the connection of the |
|
| batteries in control box |
| Check socket for pendant | |
lowering patient |
| Check pendant cable |
|
|
|
Battery Indicator shows no light |
| Check the emergency stop button is |
| released. | |
when charger is connected to |
| |
| Check charger plug for damage. | |
power supply |
| |
| Check charger cable for damage. | |
|
| |
|
|
|
Lift does not work. Battery |
|
|
indicator shows full charge. |
|
|
Notice “click” sound from control |
| Actuator problem. |
box when UP or DOWN button on |
|
|
hand control was pressed. |
|
|
|
|
|
LF2020/LF2090 | 15 |