What to do if ...
Problem | Reason | Solution | ||||||
|
|
|
|
|
|
| Batteries are flat. | Insert new batteries. |
|
|
|
|
|
|
| ||
|
|
|
|
|
|
| Memory readings can be |
|
|
|
|
|
|
|
| called up. |
|
|
|
|
|
|
|
| Measurements are not |
|
|
|
|
|
|
|
|
| |
(flashing) | possible. |
| ||||||
|
|
|
|
|
|
|
|
|
e.g. | Error codes | Refit device and repeat measurement | ||||||
................ |
| Device may not be correctly as shown and described under | ||||||
| positioned on the wrist. | «Measuring». | ||||||
|
|
|
|
|
|
|
|
|
e.g. | Error codes 30 and | Repeat measurement. If problem | ||||||
................ |
| higher: | persists, have device checked at a | |||||
| System error | Braun Service Centre. | ||||||
|
|
|
|
|
|
|
|
|
Measurement or | Right measuring position | Repeat measurement as shown and | ||||||
inflation does not | was not reached. | described under «Measuring» or de- | ||||||
start instantly. |
| activate the positioning system (A). | ||||||
|
|
|
|
|
|
|
|
|
Readings from | Blood pressure is a | none | ||||||
repeated meas- | fluctuating value. |
| ||||||
urements differ | For healthy adults, |
| ||||||
considerably. | deviations of about 10 to |
| ||||||
|
|
|
|
|
|
| 20 mmHg are possible. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Cuff is not properly fitted. | Ensure that the appliance is fitted |
|
|
|
|
|
|
|
| as follows: |
|
|
|
|
|
|
|
| – on top of the slim side of the forearm |
|
|
|
|
|
|
|
| – with the display on top |
|
|
|
|
|
|
|
| – with the battery compartment facing |
|
|
|
|
|
|
|
| the user’s body |
|
|
|
|
|
|
|
| – approx. 2 cm away from the thumb root |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Readings were not con- | For each measurement, proceed |
|
|
|
|
|
|
| sistently taken at heart | exactly as shown and described under |
|
|
|
|
|
|
| level. | «Measuring». |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Talking, coughing, | While measuring, relax, keep still, |
|
|
|
|
|
|
| laughing, moving etc. | do not move, or speak. |
|
|
|
|
|
|
| when measuring will |
|
|
|
|
|
|
|
| influence the reading. |
|
|
|
|
|
|
|
|
|
|
Prior to a mea- | Positioning system (A) is | Refer to the chapter «Select mode» | ||||||
surement, the | deactivated. | to activate the positioning system | ||||||
positioning |
| (A). | ||||||
system (A) on the |
|
| ||||||
display does not |
|
| ||||||
work. |
|
| ||||||
|
|
|
|
|
|
|
|
|
11