Bayer HealthCare Breeze 2 manual Turn the meter off, then on using

Page 46

What You See

What It Means

What You Should Do

 

 

 

 

1. You may have

1. Repeat the test and wait until

 

applied blood too

you see the “apply blood”

 

early to the test strip.

prompts before you touch the

 

 

strip to the blood drop.

 

2. There may be

2. Wick the moisture away (see

 

moisture in the test

page 34), then repeat the test.

 

strip slot.

 

 

 

 

 

There was a problem

Release the used test strip and

 

completely filling the

repeat the test with a new strip

 

test strip; accurate test

(see page 10).

 

result is not possible.

 

 

 

 

 

1. Meter has detected

1. Release the used test strip and

 

an error; accurate

repeat the test with a new strip

 

result is not possible.

(see page 10).

 

2. The 10-test disc may

2. Replace the 10-test disc, using a

 

be past its expiration

new box if necessary.

 

date or may have

 

 

been damaged due

 

 

to intense heat or

 

 

moisture exposure.

 

 

 

 

 

Your blood glucose level

Turn the meter off, then on using

 

is very high or the meter

. Retest. If the error appears

 

has detected an error;

again, call your physician or

 

accurate test result is

healthcare professional. Your

 

not possible.

blood glucose may be above

 

 

1500 mg/dL. If the problem is

 

 

with the meter, contact Bayer

 

 

Customer Service.

 

 

 

42Bayer Customer Service: 1-800-348-8100

Image 46
Contents User Guide As easy as Table of Contents Technical Information Introduction Look at Your Breeze Meter Down Buttons Explanation of Display Symbols Here is what each display element means TesTIng maTerIals Everyday use InserTIng a 10-TesT dIsCInsert Disc Counting the Number of Strips Left RunnIng a blood gluCose TesT Prepare Lancing DevicePush Out Test Strip Stick your fi nger and form a small round drop of blood Test  Bayer Customer Service Strip may not release properly Precautions Marking a Deleted Blood Test Understanding Your Test Results Expected Blood Glucose ValuesAlternative Site Testing Unusual Test ResultsRunnIng a ConTrol TesT Test Marking a Control Test Page Beyond Everyday use ReCallIng TesT resulTsViewing Your Averages Clearing Stored Test Results SeTuP funCTIon Setting the TimeSetting the Date Page Setting the Buzzer Level Setting the Time Format 12- or 24-hour Clock Setting the Date Format TransferrIng resulTs To a ComPuTer CleanIng your meTer RePlaCIng The baTTery To replace the battery, follow these steps Page Technical Information Solving Problems Troubleshooting and Error Codes What You See What It Means What You Should Do Turn the meter off, then on using This is a normal expected action Control solution with an acceptable Customer Service Replacement Items Warranty Information Bayer Customer Service Bayer HealthCare LLC Mishawaka, in 46544 USA