Graco 1750365, PD104815A owner manual Troubleshooting, Consistent low temperature readings

Page 15

Troubleshooting

1.Consistent low temperature readings.

a.The probe is not positioned properly in the ear canal. The tip of the probe must be snug and fully seated against the opening of the ear canal. Failure to properly position the probe may lead to a low temperature reading.

b.The probe lens is dirty. Clean the lens with a soft cloth moistened with water or isopropyl alcohol.

2.The word “Lo” or “Hi” appears. Temperature is out of the range of thermometer.

a.The probe is not positioned properly in the ear canal. The tip of the probe must be snug and fully seated against the opening of the ear canal. Please properly position the ear thermometer in the ear canal and then repeat the measurement.

3.Thermometer will not take a reading.

a.Check to see if the low battery icon appears with the word “Lo”. If it does, the battery needs to be replaced. See Battery Installation section of the manual.

b.Check to see if the low battery icon appears by itself. If it does, the battery may need to be replaced. See the Battery Installation section of the manual.

15

Image 15
Contents Ear Thermometer Page Do not Open EAR Thermometer For Safe Battery USE Page Page Page Overview Symbol Key Battery Installation Instructions for use See Fig Page Care and maintanance Consistent low temperature readings TroubleshootingApplied standards and approvals Page Page Page Montreal Fax Manuel Dutilisateur Page NE PAS Ouvrir LE Thermomètre Auriculaire Utilisation Sécuritaire DES Piles Page Mise EN Garde Page Aperçu Légende Installation des piles Mode d’emploi Plus consulter les figures 10 et Page Soins et entretien Dépannage Normes appliquées et approbations Page Page Page Pièces de rechange Renseignements sur la garantie au Canada