Exergen TAT-5000 manual Perspiration

Page 23

Perspiration

The TemporalScanner relies on the skin over the temporal artery to help provide an accu- rate body temperature. In fact, it is measuring the inside by measuring the outside. Your skin is a sensor, controlling body temperature in two ways: radiation and evaporation.

Since most of us don’t think about our skin as a sensor, this might be a good time to dis- cuss a little physiology.

We live our entire lives with a body temperature that changes only a few degrees. This is thanks to a very sophisticated climate control, of which the skin is a very important part. Sweating, goose bumps, and heat loss from the skin all help maintain our normal tempera- ture, keeping us comfortable.

When your internal temperature rises, your brain signals your blood to increase circulation to the skin. In this way, the body’s internal heat is carried to the surface by the blood, where it is lost by radiation.

If this is not sufficient, your sweat glands sprint into action, and perspiration is released through the pores. This liquid evaporates on your skin and you cool right down. When your temperature drops, your brain signals that heat must now be saved. Less blood circu- lates to the skin, and sweating stops.

Since there is a lot of cooling going on when you are sweating, both inside and out, it is a good idea to wait till your forehead is dry before taking your temperature with the TemporalScanner. If your forehead is sweaty, the reading would be low. Drying your fore- head could help shorten the wait, but there is another place to measure an accurate tem- perature when perspiring. It is still on the head, but in the little soft depression just behind the ear lobe, the place where young ladies are usually taught to apply perfume.

During perspiration, taking a temperature with the TemporalScanner in the area behind the ear lobe has been proven to be as accurate as a temperature taken at the temporal artery area, were it not wet. Since we sweat first on the forehead, then on the hands and feet, the chances of the area behind the ear lobe remaining dry for the measurement are excel- lent. And since we already have increased circulation to the skin during perspiration, this area will have the high blood flow necessary for the measurement.

Another instance when a high rate of blood flow on the neck can be assured is following head trauma, either surgical and accidental. At such times the neck area behind the ear lobe can be used as a primary site if the forehead is not available.

If perspiration or head trauma is not present, the area on the neck behind the ear lobe may not have sufficient blood flow to be reliable, and should not be used as the primary meas- urement site.

If there is heavy perspiration, including moisture behind the ears, wait until area is dry. For use on exercising athletes or other non-clinical subjects, contact Exergen.

Forgotten Physiology

21

Image 23
Contents TemporalScannerTM ExergenPage Important Safety Instructions Probe Cone Probe Lens Product MapOn Button Method Why the Temporal ArteryIntroduction Introduction to Temporal Artery ThermometryTable of Contents Scan Before Using, Familiarize Yourself with the InstrumentFamiliarize Yourself with the Instrument Practice Holding Your TemporalScanner Things To Know Before Taking TemperaturesUsing on an Infant Basics of Using the TemporalScanner Using the TemporalScannerUsing the TemporalScanner on a New Mother Using the TemporalScanner on a New MotherTemping Baby in Bassinette, Open Crib, or with Mom Using the TemporalScanner on an InfantFrequently Asked Questions Frequently Asked QuestionsHow does the TemporalScanner compare to our old method? Why not use only the area behind the ear lobe?What are the benefits of using temporal artery thermometry? What is arterial temperature?Conditions that could affect a reading Memorable solutions?Their solutions False low readingsWhat should I know about the instrument? Disposable Cover OptionsGeneral Rule of Thumb Guidelines for Patient Temperature AssessmentGuidelines for Patient Temperature Assessment Normal Body Temperature BTReferences Primary Points Determining a Fever ThresholdThreshold Defining Fever Threshold for Fever WorkupAn Overview of Temperature Measuring Sites Body Sites for Temperature AssessmentTympanic Membrane and Ear Temperature Reproducibility of Readings Reproducibility in Temperature MeasurementHot Blood or Cold Blood? Normal TemperatureForgotten Physiology FeverPerspiration For Kids Only Places to measure your temperatureCare and Maintenance For F/C Conversion MaintenanceCare Instructions for Fahrenheit or Celsius ConversionNormal body temperature range, otherwise Calibration Verification ProcedureCalibration Verifier Using the Portable BlackbodyPortable Blackbody Repair If repair is required Verifier SpecificationsPage Page Exergen