•Accuracy Limits: Comparison between the reference instrument(s) and the instrument being verified should be within ±0.4 ºF (0.2 ºC) for acceptable limits. If not, repeat the process. In the event they still differ by more than the acceptable limits, call Exergen for repair or replacement of the failed instrument.
Verifier Specifications:
Power Source | |
|
|
Battery Life | Approximately 1 hour continuous use. |
|
|
Low Voltage Indicator | Red LED shuts off when battery voltage drops below |
| ~5 volts. |
Temperature Range | |
|
|
Cleaning | Wipe down with alcohol or any hospital approved |
| disinfectant. Do not immerse. |
|
|
Repair
If repair is required:
•Contact Exergen at (617)
•Mark the RMA number on the outside of your package and packing slips.
•Include a description of the fault if possible.
•Send the instrument freight/postage prepaid to:
Exergen Corporation 400 Pleasant Street Watertown, MA 02472
•The instrument will be returned freight/postage prepaid.
Care and Maintenance
27