Fluke 50S, 50D warranty Calibration Connections

Page 10

50S & 50D Thermometer

Calibration Information

position until the display is locked in self-test (all LCD segments on). This thermometer is now in the Reference Junction Sensor calibration mode. Shorting out the VIEW switch grid turns off a filter so that the reading stabilizes immediately.

13.Using the 80T-150U and the Fluke 87 Series IV Digital Multimeter (DMM), measure the reference junction transistor temperature by placing the 80T-150U probe tip against Q1, located in the middle hole of the isothermal block. Wait for the temperature reading to stabilize.

14.Adjust R16 for a temperature reading that is the same as displayed on the DMM.

15.Turn the thermometer OFF by shorting the ON/OFF switch grid, and reattach the top-case.

50S or 50D

Thermometer +

_

Copper wire

5520A

zf07f.eps

Figure 5. Calibration Connections

10

Image 10
Contents Warranty ServiceIntroduction Precautions and Safety Information International SymbolsInternational Electrical Symbols Used in this Document Specifications Basic MaintenanceDisassembly/Reassembly Thermometer Disassembly and Reassembly 50D is shownChanging the Battery Replacing the Battery 50D is shownPerformance Tests Required Test EquipmentDisplay Test Wide Range Performance Test Procedure 50S Display 50D DisplayIce Bath Test Calibration Thermometer CalibrationCalibration Connections A1 Main PCA Adjustments Thermocouple Input Calibration OptionalRoom Temperature Lag Bath Qty Fluke Part Number Replaceable PartsNumber 50D Replaceable Parts DescriptionExploded View Diagram Model 50D is Shown