Fluke 50D, 50S warranty Thermocouple Input Calibration Optional, A1 Main PCA Adjustments

Page 11

Calibration

FLUKE 52-1601

Figure 6. A1 Main PCA Adjustments

zf09f.eps

Optimizing Thermometer Performance in the Negative Temperature Range

Perform the following steps to optimize the thermometer performance in the negative temperature range:

1.Put the thermometer in normal operating mode. Normal operation can be achieved by turning the thermometer OFF and then ON. No other thermometer buttons should be pressed during the power cycle operation.

2.Insert a known good thermocouple in an ice bath (see “Ice Bath Construction” earlier in this document).

3.Insert the Mercury Thermometer to the same depth as the thermocouple wires verifying the ice bath temperature is 0.0 oC ± 0.1 oC.

4.Connect the other end of the thermocouple wire to the T1 input of the thermometer and allow the instrument reading to stabilize.

5.Slowly, and in small increments, adjust R7 (T1 offset) until the thermometer reading matches the Mercury Thermometer reading. Allow the thermometer reading to stabilize and repeat the adjustment as necessary until a stable and correct reading can be attained.

6.Repeat steps 4 and 5 for R13 (T2 offset, Fluke 50D only).

Thermocouple Input Calibration (Optional)

Note

This adjustment procedure counteracts the adjustments made in the “Optimizing Performance in the Negative Temperature Range” section of this document.

This procedure optimizes the 50S or 50D for measurement with a specific thermocouple.

Perform the thermocouple input calibration using the following procedure:

1.Construct an ambient temperature lag bath as shown in Figure 7.

2.Connect the thermocouple to the thermometer input terminals.

3.Select the applicable switch postion (T1 or T2) and temperature scale (oC or oF).

4.Insert the probe into a room temperature lag bath.

11

Image 11
Contents Introduction ServiceWarranty International Electrical Symbols Used in this Document International SymbolsPrecautions and Safety Information Basic Maintenance SpecificationsThermometer Disassembly and Reassembly 50D is shown Disassembly/ReassemblyReplacing the Battery 50D is shown Changing the BatteryDisplay Test Required Test EquipmentPerformance Tests 50S Display 50D Display Wide Range Performance Test Procedure Ice Bath Test Thermometer Calibration CalibrationCalibration Connections Thermocouple Input Calibration Optional A1 Main PCA AdjustmentsRoom Temperature Lag Bath 50D Replaceable Parts Description Replaceable PartsNumber Qty Fluke Part NumberExploded View Diagram Model 50D is Shown