Fluke 50S, 50D warranty Room Temperature Lag Bath

Page 12

50S & 50D Thermometer

Calibration Information

5.Allow the system to stabilize, then adjust T1 or T2 offset adjustments for a temperature display equal to the reading of the lag bath reference thermometer.

6.Calibration of the thermometers is complete. Disconnect all test equipment from the instrument.

50S or 50D

 

Thermometer

 

+

_

 

 

Thermocouple wire

Mercury

 

 

 

 

thermometer

 

 

(0.02°C resolution)

 

 

Lag bath

 

 

(room temperature

 

 

water)

 

 

Immersed to

 

 

approximately

 

 

same depth

 

 

zf06f.eps

Figure 7. Room Temperature Lag Bath

12

Image 12
Contents Service WarrantyIntroduction International Symbols Precautions and Safety InformationInternational Electrical Symbols Used in this Document Specifications Basic MaintenanceDisassembly/Reassembly Thermometer Disassembly and Reassembly 50D is shownChanging the Battery Replacing the Battery 50D is shownRequired Test Equipment Performance TestsDisplay Test Wide Range Performance Test Procedure 50S Display 50D DisplayIce Bath Test Calibration Thermometer CalibrationCalibration Connections A1 Main PCA Adjustments Thermocouple Input Calibration OptionalRoom Temperature Lag Bath Replaceable Parts NumberQty Fluke Part Number 50D Replaceable Parts DescriptionExploded View Diagram Model 50D is Shown