Omega Speaker Systems HH303 manual Calibration Procedure

Page 14

Calibration Procedure

Model No. HH302 & HH303

Environmental: 23 ± 5

Date

JUN./06/2002

 

 

Page

1 of 1

K,J Thermometer (HH302) calibrate procedure

1.Hold REL+AVG+TIMER then press power , release all button ,press HOLD +C/F in 5 seconds, LCD will show “CAL 1”.

2.Input 52.418 mV via copper wire , press REL button when LCD display a stable value(about 14,000 count). Then LCD will display “CAL 2”

3.Input 33.290 mV via copper wire, press REL button when LCD display a stable value(about 18,000 count), Then LCD will display “CAL 3”

4.Input 0 C via thermocouple wire, press REL button when LCD display a stable

value(less then 1000 count),

Now , press REL button again to finish calibrating .

K,J Thermometer (HH303) calibrate procedure

1.Hold REL+AVG+ T1-T2 then press power , release all button ,press HOLD +C/F in 5 seconds, LCD will show “CAL 1”.

2.Input 52.418 mV via copper wire , press REL button when LCD display a stable value(about 14,000 count). Then LCD will display “CAL 2”

3.Input 33.290 mV via copper wire, press REL button when LCD display a stable value(about 18,000 count), Then LCD will display “CAL 3”

4.Input 0 C via thermocouple wire, press REL button when LCD display a stable value(less then 1000 count),

Now , press REL button again to finish calibrating .

Omega

Image 14
Contents Temperature PH/CONDUCTIVITYOmega.comTM III. Symbol Definition and Button IV. Operation InstructionsII. Specifications IntroductionAccessory Temperature CoefficientDimension 184 64 30mm Power requirementIII. Symbol Definition and Button Location $34$556 $/!6 IV. Operation Instructions MAX/MIN/AVG Operation for Main Display Auto Power OffLow Battery Condition Command M Command KCommand T RS232 command Function RemarksCommand a Command RCommand C Represent ###.# Byte first two BCD code of main window value Calibration Procedure