Omega Vehicle Security HH502 manual HH 502/HH 503 Calibration Procedure

Page 12

HH 502/HH 503 CALIBRATION PROCEDURE

Note: The following calibration procedure should perform only by qualified technicians who have access to the items as following.

Equipment: The class of calibrator had better 10 times greater than the measured meter.

1.Set the jumper position on J1 position.

2.The input selection change to T1 position, then input DCV 60mV after the display is stabilized and press “ ENTER “ key.

3.Input 0￿F after the display is stabilized and press “ ENTER “ key to complete calibration.

4.After adjustment set the jumper back to J3 position.

1

Image 12
Contents Omega Omega.comTM Safety Information IntroductionSpecifications ElectricalEnvironmental Power Switch F Selecting the Temperature ScaleHold Mode only Main display Operating InstructionsMIN MAX with Time record Mode only Main display T1 /T2/T1-T2 Main display Input SelectionJ T1 Input Thermocouple Type Select Main display REL Relative mode only Main display Sec. Minu. Selecting the Time scale 12 Hi/Lo Limits mode only Main display 10 T1/T2/T1-T2 second display Input Selection11 K/J T2 Input thermocouple type select second display Operator Maintenance Battery ReplacementHH 502/HH 503 Calibration Procedure Warranty / Disclaimer Temperature