Summer Infant 03170 manual Troubleshooting

Page 10

TROUBLESHOOTING (continued)

 

Conditions

Possible cause and solution

 

 

 

sign is displayed

• Low battery signal. Replace

 

 

with new battery.

 

 

• Measurement cannot be made

• Tip of probe may be dirty or

 

or abnormal value is

damaged. Clean the probe.

 

displayed.

• Excessive earwax in the ear

 

 

canal. Clean the ear canals

 

 

with a cotton swab.

 

 

• Probe may be insterted

 

 

incorrectly. Insert the

 

 

probe correctly.

 

 

Follow inserting the probe into

 

 

the ear canal under “use of

 

 

your product” page 8.

 

 

If the problem still persists,

 

 

send to service center

 

 

for repair.

 

 

 

is displayed

• There may be a short circuit.

 

 

Please contact service center.

 

 

• The ambient temperature has

 

 

exceeded the specific range

 

 

of 41°F-104°F(5°C-40°C).

 

 

• Measurement will not start

• Measurement is made when

 

after pressing the button.

the “ “ is not displayed.

 

 

Measure again after the sign

 

 

is displayed.

 

 

• Measured result is “Hi” or “Lo”

• ”Hi” is displayed when the

 

 

body temperature is more

 

 

than 107.6°F(42°C).

 

 

• ”Lo” is displayed when the

 

 

body temperature is less

 

 

than 95°F(35°C).

 

 

 

10

Image 10
Contents ONE Second EAR Thermometer Contents SafetyNormal body temperature Why Measure in the Ear?How Does the Ear Thermometer Work? Parts and Accesories Index Features and EffectivenessFeatures Main BodyGeneral Preparation Before USEInserting batteries Preparation Before USE Correct Holding PosturePower on USE of Your ProductUSE of Your Product Conditions Possible cause and solution TroubleshootingPower off Starting the measurementTroubleshooting Probe Product Care and StorageBatteries StorageSummer Infant, Inc