CHAPTER 4
SPECIFICATIONS
4.1 AGRI-THERM II (Model 6110.4 ZL) SPECIFICATIONS
TEMPERATURE MEASUREMENT
Scale Range: | |
Resolution: | 0.1˚C or 0.1˚F |
Accuracy: | ±0.5˚C or ±1.0˚F |
Repeatability: | ±0.1˚C or ±0.2˚F |
Temperature: | All Functions in ˚C or ˚F, corresponding to voltage out |
Noise Effective Temperature: | <0.1˚C or ºF |
Measurement Modes: | Surface Temperature, Dry Bulb Temperature and |
| Temperature Differential (Patent No. 4,301,682) |
OPTICAL CONSIDERATIONS
Optical Configuration: | Robust, |
| Optics per Military Specification |
Spectral Pass Band: | 8 < Wavelengths < 14 Microns |
Sighting: | |
| Visible Light Illuminates Field of View (FOV) |
Illumination Source: | Light Emitting Diode |
Focus: | Variable from 2º to 20º with |
| Pending) |
Display: | Numeric on Liquid Crystal Display on Rear Panel |
OPERATING CONDITIONS
Operating Environment: | |
| Humidity, |
Storage Temperature: | Same as Operating Environment Temperature |
GENERAL
Response Time: | <1 Second |
Emissivity Compensation: | Settable from 10% to 99% on Rear Panel |
Operating Distance: | 2 cm to Infinity |
ELECTRICAL INTERFACE
Power Source: | Rechargeable Nickel Cadmium Rechargeable Batteries |
| 200 Hours Continuous Use on Full Battery Charge |
Output Signal: | |
Battery Charger: |
WARRANTY
Warranty:
One Year-Limited Warranty of Parts & Labor
Everest Interscience, Inc. has a policy of continued design upgrade. Therefore, we reserve the right to change specifications without notice
10