Control Solution Results

When the control test result falls within the control range, the meter has passed the QC check and can be used to test patient blood.

If the result falls outside the expected range, repeat the test. Results that fall outside the expected range may indicate:

procedural error

control solution is expired, contaminated, beyond its discard date, or inadequately mixed

code number in meter does not match code number on test strip bottle

dirty meter or test strip holder

problem with test strip (expired, damaged, etc)

control solution testing temperature outside the range of 18°–30°C (64°–86°F)

too much or too little control solution applied to test strip

28 C h a p t e r 2

Image 38
Contents Hospital Meter Operator’s Guide Page Page 2002, 1997 LifeScan, Inc. All rights reserved Table of Contents Troubleshooting Settings and MemoryCleaning and Maintenance Preface About this Guide Conventions Used in this Guide Help 888 353-0800 CanadaIntroduction Introduction Learning the SureStep MeterClosed Opened Display Symbols Meter Memory Result Apply Blood Sample to the Test Strip Meter displays results between 0-500 mg/dL 0-27.8 mmol/L HI Result Error Messages SureStepPro Test Strips Front Back H a p t e r T r o d u c t i o n Precautions T r o d u c t i o n Infection Control Operation Coding the MeterOperation Match the code numbers If the code numbers do not match, code the meterQuality Control Test Glucose Control Solution Operation Check the Code Performing a Quality Control TestCha pt e r E r a t i o n H a p t e r E r a t i o n Control Solution Results Linearity Test Patient Blood Test Performing a Patient Blood Test Obtain a drop of blood Apply blood to the test strip as followsSyringe Front Patches or streaks H a p t e r E r a t i o n Patient Results Settings and Memory Meter SettingsSettings and Memory Display of Last Result Glucose Units Beep signals on Using the Meter for the First Time Meter MemoryRecalling All Results in Meter Memory Order of test result Recalling Most Recent Result After Testing Settings and Memory Cleaning Maintenance General CareCleaning and Maintenance CleaningCleaning the Outside of the Meter Clean the test strip holder Remove the test strip holderE a n i n g a n d M a i n t e n a n c e Replace the test strip holder Turn on the meter H a p t e r Replacing the Batteries Turn off the meter and turn it overRemove the old batteries Insert new batteries Replace the battery doorTurn on the meter to verify power Cleaning and Maintenance Troubleshooting Troubleshooting Error Messages What To DoLow Test strip Check test strip Test strip Obtain a laboratory Result, Area, but Temperature Contact LifeScan for help Electronics Symbol Meaning What To Do Helpful MessagesIf symbol is Specifications Hematocrit Range Index Error messages 9 Page Page AW 050-097-03A LifeScan, Inc 11/02
Related manuals
Manual 74 pages 3.73 Kb