TROUBLESHOOTING
Warning: If you experience a problem with your lift and are unable to service it yourself, contact your Lumex dealer.
LF1030 HYDRAULIC PATIENT LIFT
SYMPTOM | POSSIBLE CAUSE | ACTION |
Patient lift does not | Air in the cylinder | Bleed the cylinder by turning the release knob |
go up or down |
| and push down on the mast. Pump up the lift |
| Loss of hydraulic fluid | again and turn the release knob again. |
| (leaking cylinder) |
|
|
| Replace any leaking cylinder |
Patient lift does not | Air in the cylinder | See above Directions |
go down |
|
|
Note: If the hydraulic cylinder begins leaking fluid, do not operate. Contact your Lumex dealer for a new cylinder.
LF1040 ELECTRIC PATIENT LIFT
SYMPTOM | POSSIBLE CAUSE | ACTION |
Control box is not | Battery discharged | Charge battery |
responding and relay | Battery charger defective | Replace battery charger |
is not clicking | Battery defective | Replace battery |
| Hand control defective | Replace hand control |
Actuator does not run, | Actuator plug not properly | Push actuator plug firmly into control box |
but relay is clicking | seated |
|
| Actuator defective | Replace actuator |
| Control box defective | Replace control box |
Actuator goes up or | Hand control defective | Replace hand control |
down, but not both | Control box defective | Replace control box |
LF1030 HYDRAULIC PATIENT LIFT / LF1040 ELECTRIC PATIENT LIFT OWNER'S MANUAL, 10/03 | 21 |