Monarch 110 DC Voltage Measurements, AC/DC Current Measurements, Resistance Measurements

Page 10

MODEL 110 AUTORANGING DIGITAL MULTIMETER

5.2 DC Voltage Measurements

WARNING

Maximum Input Voltage is 600VDC. To avoid electrical shock hazard and/or damage to this instrument, do not attempt to make any voltage measurement that may exceed this limit.

1. Connect the red test lead to the ‘V’ jack and the black test lead to the ‘COM’ jack.

2. Set the Function Selector switch to the V range.

3. Connect the test leads in parallel with the circuit being measured.

4. Read the measured voltage on the instrument display.

5.3 AC/DC Current Measurements

WARNING

To avoid injury, do not attempt a current measurement if the open circuit voltage exceeds the rated voltage of the meter.

1.If the current to be measured is unknown, connect the red test lead to the ‘A’ jack. If it is known or has been determined that the current is 400mA or less, connect the red lead to the ‘µA/mA’ jack. Connect the black test lead to the ‘COM’ jack.

2.Set the Function Selector switch to the current range corresponding to the jack being used and press the blue button to select AC or DC.

3.Remove power from the circuit to be measured and connect the instrument in series with this circuit. Connect the black lead to the negative (-) side and the red lead on the positive (+) side being measured.

4.Apply power to the circuit and read the measured current on the instrument display.

7

MODEL 110 AUTORANGING DIGITAL MULTIMETER

uA or mA measurement

Amp measurement

5.4 Resistance Measurements

WARNING

Before making any in-circuit measurements, remove power from the circuit being tested and discharge all capacitors in the circuit.

1.

Connect the red test lead the to :jack

 

and the black test lead to the ‘COM’ jack.

2.

Set the

Function

Selector

switch to

 

the

range

and press

the blue

 

button to select the : function.

 

3.

Remove power from the circuit being

 

tested and discharge all capacitors in

 

the circuit.

 

 

4.

Connect the test leads in parallel with

 

the circuit to be measured and read the

measured resistance on the instrument display.

8

Image 10 Contents
Monarch Autoranging Digital MultimeterSafeguards and Precautions Limited WarrantyTable of Contents Declaration of ConformitySpecifications IntroductionContinuity Beeper Range DC Current uA and mA AutorangingDiode Tester Battery life Auto Power Off Power requirement SafetyReplacing Fuses Replacing BatteriesΜA, mA, Hz Control Functions Continuity Measurements Low Battery ConditionOperating Instructions Fuse ProtectionAC/DC Current Measurements DC Voltage MeasurementsResistance Measurements