Omega Vehicle Security CCT-05 Series manual Changing the Signal Output

Page 6

SIGNAL OUTPUT module MS

The signal conditioner provides two different analog output signals, both proportional to the signal input.

Output in Current : 4 to 20 mA, terminals 9 - 10

Output in Voltage : 0 to 10 Vdc, terminals 11 - 12

Do not use both outputs simultaneously. Only one selection can be made.

The side label shows which one is selected.

1

3

4

5

6

H i

 

 

 

 

/ ALIMENTACION

INPUT / ENTRADA

 

 

 

 

SUPPLY

OUTPUTS / SALIDAS

 

mA

Vdc

+

-

+

-

Lo

7

9

10

11

12

CHANGING THE SIGNAL OUTPUT

FIG.5

All signal conditioners are delivered as a standard version, with the analog output selected as 4 to 20 mA, unless specified otherwise. To select a 0 to 10 V output, remove jumpers E and F as shown in Figure 6. Other non-standard output voltage and current ranges may be obtained by adding and/or replacing resistors given in Tables 2 and 3.

NON STANDARD OUTPUT

 

NON STANDARD OUTPUT

VOLTAGES

TABLE 2

 

 

CURRENTS

TABLE 3

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Output

 

Value in KΩ

for :

 

 

Output

Value in Ω for :

 

in V.

R29

 

R30

R31

R32

 

in mA.

R18

R24

R25

 

 

 

 

 

 

 

 

 

 

 

±10

49.9

 

-----

200

----

 

0 to 5

-----

100

-----

0 to 1

-----

 

-----

11

100

 

0 to 10

-----

49.9

-----

0 to 5

-----

 

-----

100

100

 

1 to 5

100 K

124

-----

1 to 5

-----

 

100

66.5

100

 

0 to 20

-----

-------

24.9

 

 

 

 

 

 

 

 

 

 

 

 

 

"-----" means "Resistor must not be installed"

 

E F Jumpers E and F : Closed only if the output is 4 to 20 mA.

 

Replace or add the indicated

R18

resistors with the values shown

in Tables 2 and 3 for the desired

 

 

output.

R29

 

R30

 

 

FIG.6

R24

R25

R31

R32

5

Image 6
Contents User’sGuide Omega.comTM Table of Contents Procedure for Disassembly Open the HousingProcedure Internal OverviewChanging the Power Supply Recomended WiringPrecautions ProtectionsChanging the Signal Output NON Standard Output Voltages CurrentsOffset Adjust OverviewAmplifier Gain Sensor Type Selection CCT-05 Electrical FeaturesGeneral Specifications InputSignal Input Range Selection Adjustment and Calibration ProcedureSignal Output ConnectionsDimensions Technical DataMechanical Page USAWARRANTY/DISCLAIMER Temperature